I
NSTALLATION
- 18 -
4-13
H
OW TO
S
HOW
R
ESISTIVITY OR
C
ONDUCTIVITY
U
NITS
(C01)
W
ITH
T
EMPERATURE
C
OMPENSATED OR NON
T
EMPERATURE
C
OMPENSATED VALUES
Temperature compensation is a way of ‘standardizing’ Resistivity or Conductivity to measurements
that would be seen if the water temperature was 25
C. The Compensation Mode is enabled by
default.
1.
Press the Main and
“+“ Buttons together
to enter the menu.
The Display will
show C01 and the
units chosen. The
following Display
shows Temperature
Compensated
Resistivity Units:
M
.cm @25
C.
2.
Press the “+” or “-“
Button to select
Non Temperature
Compensated
Resistivity Units:
M
.cm.
3.
Press the “+” or “-“
Button to select
Temperature
Compensated
Conductivity Units:
S/cm @25
C.