I
NSTALLATION
- 16 -
4-10
P
URGING
A
IR FROM THE
S
YSTEM
At this time you should have installed the
SimpliPak Cartridge, Barbed Fitting and the
Tygon Tubing. Air trapped in the SimpliPak
Cartridge should now be purged from the system.
Press the Main Button to put the system into
DISPENSING mode (A).
Dispense 2 Litres of water from the system.
Stop dispensing by pressing on the Main Button.
4-11
H
YDRATING THE
S
YSTEM
At this time, the Final Filter is not installed.
Leave the system overnight or for several hours
in PRE OPERATE mode (see
Section 5-4
Operating Modes
). The system will regularly
recirculate water and rinse off purification media
inside the SimpliPak Cartridge. Do not leave the
system in STANDBY mode.
(The next morning) Press the Main Button to put
the system into DISPENSING mode.
Allow about 0.5 Litre of water to be dispensed
from the system.
Press the Main Button. The system will stop
dispensing water.
Remove the Tygon Tubing and the Barbed Fitting
from the POU Dispenser.
The Final Filter needs to be installed. See
Section
Installation and Rinsing of the Final Filter.
A