Millipore Amicon Ultra-2 User Manual Download Page 4

Amicon

®

 Ultra-2 Centrifugal Filter Devices

4

Performance - Protein Concentration

Flow Rate

Factors affecting flow rate include sample concentration, starting vol ume, chemical nature of solute, relative centrifugal force, 

centrifuge rotor angle, membrane type, and temperature. Figures 1 and 2 and Tables 2 and 3 can be used to estimate the time 

required to achieve a given volume of filtrate or concentrate for a variety of protein markers. A typical spin time for a 2 mL sample 

in a fixed angle rotor is approximately 10 to 60 minutes (depending on device nominal molecular weight limit). While most of 

the sample is filtered in the first 10 to 20 minutes of centrifugation, the lowest concentrate volume (30–70 μL) is reached after 

spinning for 10 to 60 minutes.

Figure 1.  Typical Filtrate Volume vs. Spin Time for Amicon

®

 Ultra-2 Device, Swinging Bucket Rotor

0

0.5

1

1.5

2

0

10

20

30

40

50

60

3K
10K
30K
50K
100K

Filtr

ate 

Volume 

(mL)

Spin time (min)

0

0.5

1

1.5

2

0

10

20

30

40

50

60

3K
10K
30K
50K
100K

Spin time (min)

Filtr

ate 

Volume 

(mL)

Spin conditions: Swinging bucket rotor, 4,000 × g, room temperature, 2 mL starting volume. 

 

Protein markers used: Cytochrome c for 3K and 10K, BSA for 30K and 50K, and IgG for 100K, n=8.

Figure 2.  Typical Filtrate Volume vs. Spin Time for Amicon

®

 Ultra-2 Device, Fixed Angle Rotor

Spin conditions: 35° fixed angle rotor, 7,500 × g, room temperature, 2 mL starting volume. 

 

Protein markers used: Cytochrome c for 3K and 10K, BSA for 30K and 50K, and IgG for 100K, n=8.

Summary of Contents for Amicon Ultra-2

Page 1: ...pun in a swinging bucket or fixed angle rotor Amicon Ultra 2 devices are supplied non sterile and are for single use only The Amicon Ultra 2 product line includes 5 different cutoffs Molecular Weight...

Page 2: ...Failure to fully seat the device in the filtrate collection tube and push the concentrate collection tube firmly onto the device may result in the device breaking during centrifugation See figure bel...

Page 3: ...mL protein in 100 mM NaCl 50 L of 40 mg mL protein in 100 mM NaCl 100 mM NaCl 50 L of 40 mg mL protein in 12 3 mM NaCl 12 3 mM NaCl Add 1 95 mL of 10 mM NaCl or exchange buffer 2 mL of 1 mg mL protei...

Page 4: ...10 to 20 minutes of centrifugation the lowest concentrate volume 30 70 L is reached after spinning for 10 to 60 minutes Figure 1 Typical Filtrate Volume vs Spin Time for Amicon Ultra 2 Device Swingin...

Page 5: ...nd 10K BSA for 30K and 50K and IgG for 100K n 8 Shaded volumes were used for the calculation of protein recovery in Table 5 Table 3 Typical Concentrate Volume Concentration Factor v Spin Time Fixed An...

Page 6: ...nging bucket rotor 4 000 g or 35 fixed angle rotor 7 500 g 2 mL starting volume room temperature n 12 Factors that determine sample recovery include the nature of the protein solute relative to the de...

Page 7: ...te collection tube and concentrate collection tube before use 2 Fill filter device with solution and reweigh 3 Assemble device in filtrate collection tube and centrifuge per instructions 4 Collect the...

Page 8: ...rum or plasma do not exceed 5 400 x g Active membrane area 1 cm2 Hold up volume 5 L Dimensions Filter device and tube Length concentration mode device in filtrate tube 119 7 mm 4 71 in Length recovery...

Page 9: ...ent 0 1 Tergazyme detergent 1 Lubrol PX detergent 0 1 Triton X 100 surfactant 0 1 Nonidet P 40 surfactant 2 Tween 20 surfactant 0 1 Sodium deoxycholate 5 Organic solvents Acetone Not recommended Ethyl...

Page 10: ...FC900396 10K 8 24 96 500 UFC501008 UFC501024 UFC501096 UFC5010BK UFC201024 UFC801008 UFC801024 UFC801096 UFC901008 UFC901024 UFC901096 10K IVD 8 24 96 UFC801008D UFC801024D UFC801096D UFC901008D UFC90...

Page 11: ...office nearest you go to www sigmaaldrich com offices Technical Assistance Visit the tech service page on our web site at www sigmaaldrich com techservice Standard Warranty The applicable warranty fo...

Reviews: