Millipore Amicon Ultra-2 User Manual Download Page 1

Amicon

®

 Ultra-2 Centrifugal Filter Devices

1

User Guide

Amicon

®

 Ultra-2 

Centrifugal Filter Devices

for volumes up to 2 mL

For research use only;

not for use in diagnostic procedures.

Introduction

Amicon

®

 Ultra-2 centrifugal filter devices provide fast ultrafiltration, with the capability for high concentration factors and easy 

concentrate recovery from dilute and complex sample matrices. The vertical design and available membrane surface area provide 

fast sample processing, high sample recovery (typically greater than 90% of dilute starting solution), and the capability for 

 

50-fold concentration. Typical processing time is 10 to 60 minutes depending on Molecular Weight Cut Off (MWCO). Solute 

polarization and subsequent fouling of the membrane are minimized by the vertical design, and a physical deadstop in the filter 

device prevents spinning to dryness and potential sample loss. Efficient recovery of the concentrated sample (retained species) is 

achieved by a convenient reverse spin step after collecting the filtrate. The device can be spun in a swinging bucket or fixed angle 

rotor. Amicon

®

 Ultra-2 devices are supplied non-sterile and are for single use only.

The Amicon

®

 Ultra-2 product line includes 5 different cutoffs (Molecular Weight Cut Off, MWCO). These devices are for 

 

research use only and not for use in diagnostic procedures.

 

• Amicon

®

 Ultra 3K device — 3,000 MWCO

 

• Amicon

®

 Ultra 10K device — 10,000 MWCO

 

• Amicon

®

 Ultra 30K device — 30,000 MWCO

 

• Amicon

®

 Ultra 50K device — 50,000 MWCO

 

• Amicon

®

 Ultra 100K device — 100,000 MWCO

Applications

 

Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids (DNA/RNA samples, either 

 

single- or double-stranded), microorganisms, column eluates, and purified samples

 

Purification of macromolecular components found in tissue culture extracts and cell lysates, removal of primer, linkers, or 

 

molecular labels from a reaction mix, and protein removal prior to HPLC

 

Desalting, buffer exchange, or diafiltration

Materials Supplied

10K

4

3

2

1

00

20

00

00

00

Filter Device

Concentrate  

collection  

tube

Filtrate  

collection  

tube

The Amicon

®

 Ultra-2 device is supplied with two tubes. During 

operation, one tube is used to collect filtrate; the other to cap 

the device during concentration and subsequently to recover the 

concentrated sample.

Summary of Contents for Amicon Ultra-2

Page 1: ...pun in a swinging bucket or fixed angle rotor Amicon Ultra 2 devices are supplied non sterile and are for single use only The Amicon Ultra 2 product line includes 5 different cutoffs Molecular Weight...

Page 2: ...Failure to fully seat the device in the filtrate collection tube and push the concentrate collection tube firmly onto the device may result in the device breaking during centrifugation See figure bel...

Page 3: ...mL protein in 100 mM NaCl 50 L of 40 mg mL protein in 100 mM NaCl 100 mM NaCl 50 L of 40 mg mL protein in 12 3 mM NaCl 12 3 mM NaCl Add 1 95 mL of 10 mM NaCl or exchange buffer 2 mL of 1 mg mL protei...

Page 4: ...10 to 20 minutes of centrifugation the lowest concentrate volume 30 70 L is reached after spinning for 10 to 60 minutes Figure 1 Typical Filtrate Volume vs Spin Time for Amicon Ultra 2 Device Swingin...

Page 5: ...nd 10K BSA for 30K and 50K and IgG for 100K n 8 Shaded volumes were used for the calculation of protein recovery in Table 5 Table 3 Typical Concentrate Volume Concentration Factor v Spin Time Fixed An...

Page 6: ...nging bucket rotor 4 000 g or 35 fixed angle rotor 7 500 g 2 mL starting volume room temperature n 12 Factors that determine sample recovery include the nature of the protein solute relative to the de...

Page 7: ...te collection tube and concentrate collection tube before use 2 Fill filter device with solution and reweigh 3 Assemble device in filtrate collection tube and centrifuge per instructions 4 Collect the...

Page 8: ...rum or plasma do not exceed 5 400 x g Active membrane area 1 cm2 Hold up volume 5 L Dimensions Filter device and tube Length concentration mode device in filtrate tube 119 7 mm 4 71 in Length recovery...

Page 9: ...ent 0 1 Tergazyme detergent 1 Lubrol PX detergent 0 1 Triton X 100 surfactant 0 1 Nonidet P 40 surfactant 2 Tween 20 surfactant 0 1 Sodium deoxycholate 5 Organic solvents Acetone Not recommended Ethyl...

Page 10: ...FC900396 10K 8 24 96 500 UFC501008 UFC501024 UFC501096 UFC5010BK UFC201024 UFC801008 UFC801024 UFC801096 UFC901008 UFC901024 UFC901096 10K IVD 8 24 96 UFC801008D UFC801024D UFC801096D UFC901008D UFC90...

Page 11: ...office nearest you go to www sigmaaldrich com offices Technical Assistance Visit the tech service page on our web site at www sigmaaldrich com techservice Standard Warranty The applicable warranty fo...

Reviews: