Weight Specifications
The location where the Water Purification System is installed will need to fully support its
Operating Weight:
System type
Dry weight (Kg/Lb)
Shipping weight (Kg/Lb)
Milli-Q
®
HX 7040
78 / 172
97 / 213
Milli-Q
®
HX 7080
86 / 190
105 / 231
Milli-Q
®
HX 7120
94 / 208
113 / 249
Milli-Q
®
HX 7150
105 / 232
124 / 273
Milli-Q
®
HR 7060
72 / 166
91 / 207
Milli-Q
®
HR 7120
75 / 179
94 / 220
Milli-Q
®
HR 7170
78 / 192
97 / 233
Milli-Q
®
HR 7220
84 / 221
103 / 252
Shipping Weight is defined as a dry system in its shipping container. It does not include
the consumable or any accessories.
Dry Weight is defined as a system without its shipping container. It does not include the
consumable or any accessories.
The UF pre filtration option weight is 22 kg / 121.3 lb
Electrical Specifications
The systems can be supplied by Mains electricity 90-253 VAC with a frequency range from
48 to 62Hz.
System type
Voltage
Power Consumption (VA)
Milli-Q
®
HX 7040/7080
220-240 VAC @ 50/60 Hz
750
Milli-Q
®
HX 7120/7150
220-240 VAC @ 50/60 Hz
870
Milli-Q
®
HX 7040/7080
120 VAC @ 60 Hz
775
Milli-Q
®
HX 7120/7150
120 VAC @ 60 Hz
900
Milli-Q
®
HX 7040/7080
100 VAC @ 50/60 Hz
775
Milli-Q
®
HX 7120/7150
100 VAC @ 50/60 Hz
900
Milli-Q
®
HR 7060/7120
220-240 VAC @ 50/60 Hz
620
Milli-Q
®
HR 7170/7220
220-240 VAC @ 50/60 Hz
750
Milli-Q
®
HR 7060/7120
120 VAC @ 60 Hz
620
Milli-Q
®
HR 7170/7220
120 VAC @ 60 Hz
750
Milli-Q
®
HR 7060/7120
100 VAC @ 50/60 Hz
620
Milli-Q
®
HR7170/7220
100 VAC @ 50/60 Hz
750
The source of electrical power must be earth grounded.
Specifications | 13