Consumables Application provides information related to your
system's consumables.
Progard
®
pack: Lists the Progard
®
pack(s) details. The catalog number, installation date,
processed volume and replacement date.
RO cartridge: Identifies your RO cartridge using type, lot and catalog number. The
installation date can also be checked here.
Vent filter: Lists the tank vent filter type, lot and catalog numbers with the installation and
replacement dates.
Loop filter: Lists your Loop filter details. The catalog number, installation date, processed
volume and replacement date.
Measures Application gives information on the measurements from
your system.
Water quality measurements:
provides information about each stage of the water
purification process in the system.
• Tap water: Conductivity and temperature are monitored.
• RO stage: Conductivity, temperature and the permeate and ionic rejection outputs are
monitored.
• ELIX
®
stage: On HX systems, the Elix
®
product conductivity and temperature are
monitored.
• Distribution stage: If the kits are installed, TOC, Distribution Resistivity and
temperature are displayed.
Hydraulic measurements & actuator statuses:
shows all hydraulic measurements
and gives the status of each actuator.
• Tap water: The tap feed pressure and if applicable, the external solenoid valve status
is displayed.
• RO stage: All parameters related to the RO stage are displayed. For example, the RO
pressure, flowrate and recirculation and the different solenoid valve statuses.
• Elix
®
stage: On HX systems, the parameters related to the Elix
®
stage are displayed.
In this application, the Elix
®
status, product flowrate and recovery, and the UV lamp
status are monitored. TOC, distribution resistivity and temperature are also monitored if
the options are installed in the system.
• Storage stage: The tank level, and, if applicable, the ASM UV lamp status are
displayed.
• Distribution stage: If applicable, the distribution pump status and the dumping valve
status are displayed.
Electrical measurements:
shows electrical measurements.
System Display | 27