3. Remove the filter from its housing, leveraging with the finger.
4. Insert the new filter into the housing, being sure that the filter direction (see the blue arrow on the
filter) corresponds with the black arrow on the unit.
5. Now insert the metal part of the long hose into the cover hole then connect the other filter hose to its
place.
After changing the filter, reset the defrost pump hour counter.
Click on the icon
from the Settings page. It is necessary to insert the Administrator password
(issued as a separate document inserted in the Operator Manual). Then press YES to the below question: