Personnel:
n
Trainer
1.
Press the multi-function button.
ð
The current workout stops.
2.
Turn off the device via the on/off switch on its underside
ð
The network connection is interrupted.
3.
Disconnect the cable connections from the athlete/patient.
4.
Initiate first aid measures.
Stop button
The multi-function button is not only used for choosing and con-
firming the programs, but also as a stop button. The training can be
stopped at any time with the multi-function button.
Turning up the intensity during the pulse pause is limited to 5 units
both on the level controllers and the main controller.
Main controller
If the “Load level” function is enabled, the device is programmed
such that the main controller cuts out at 30, 40, and 50 for safety
reasons. Regardless of whether or not the “Load level” function is
enabled, an additional cut-out is set for all programs and training
plans when the device reaches a value of 85. Before the athlete
can increase the intensity beyond this value, the main controller
must be acknowledged by pressing it once.
Personnel:
n
Athlete/Patient
n
Trainer
The training session can be interrupted between individual exer-
cises to pause or discuss the next exercise with the trainer.
1.
Press and briefly hold the main controller (Fig. 3/14).
ð
The main level jumps to “0” and all LEDs are flashing.
The training time continues running during the pause.
2.
Turn the main controller to end the pause.
ð
The training session continues as usual.
Fig. 56: On/off switch
Stop button
Turning up the intensity
Pausing the training session
Overview of programs/training plans
2021-07-28
miha-bodytec II EMS training device
93