Application technology
34
Program overview
At delivery, the reprocessing programs are assigned to the A 503 mobile unit. The follow-
ing modules can be used in various combinations:
- A 612
- A 620
- A 621
- A 622
Exceptions to this are the
Vials
and
Pipettes
programs (see table).
In order to use other mobile units, new load carriers must be set up in the web interface
under “Load carrier management”. “Load carrier management” is described in the pro-
gramming manual.
Power levels
The mobile units in the modular loading concept can be fitted with modules in different
ways. Depending on the type and number of modules used, different volumes of water and
pump speeds are required to achieve a similar wash pressure. The individual reprocessing
programs are therefore available with different power levels. The higher the power level of a
reprocessing program, the higher the volume of water and the pump speed.
The program structure and the following parameters, which are relevant to performance,
are identical in the different power levels:
- Water quality
- Temperature
- Holding time
- Dispensed process chemical in %
- Drying temperature
- Drying duration
(see appendix: “Power levels/mobile units/module combinations” table)