Program settings
88
Adjusting program settings
The program settings should be adjusted to suit technical require-
ments and the wash items.
Additional specialist knowledge is required to alter program settings
and this should therefore be undertaken only by experienced users or
by Miele Service.
Program and dispensing changes must be documented for vali-
dated processes according to local regulations.
The processes must be validated again as necessary.
Program structure
Each program is subdivided into program blocks which run one after
another. A program consists of at least one and a maximum of 11
program blocks. Each block can occur only once in a program.
The so-called program header is placed above the program blocks
and contains general program settings. Individual wash block param-
eters are also globally activated or deactivated here.
Program header
-
Water volume change
Every program block with water intake is allocated a nominal water
quantity. The water quantity can be raised or lowered incrementally
for all blocks to the base value in the program chart.
-
Drainage time
If the on-site waste water system is insufficient to drain the waste
water from the wash cabinet within the time allocated, the drainage
time can be lengthened by a set amount.
Parameters for measuring wash pressure and spray arm monitoring
can only be accessed by Miele Service.