![Miele PG 8581 Operating Instructions Manual Download Page 137](http://html.mh-extra.com/html/miele/pg-8581/pg-8581_operating-instructions-manual_234475137.webp)
Technical details
137
Imperial
Metric
Height with machine lid
Height without machine lid
32 7/8"
32 5/16"
835
mm
820
mm
Width
23 9/16"
598
mm
Depth
Depth with door open
23 9/16"
47 1/4"
598
mm
1,200
mm
Wash cabinet dimensions:
Height
Width
Depth of upper basket/lower basket
20 9/16"
21 1/8"
20 9/16"
520
mm
530
mm
474
mm/
520
mm
Weight (net)
163
lbs
74
kg
Max. load capacity of open door
81.6
lbs
37
kg
Voltage, rated load, fuse rating
See data tag
See data tag
Power cord
Approx.
5' 9"
ft.
Approx.
1.8
m
Water connection temperature:
Cold water/steam condenser
Hot water (optional)/DI water (optional)
max.
68
°F
max.
150
°F
Max.
20
°C
Max.
65
°C
Static water pressure
max.
145
psi
Max.
1,000
kPa pressure
Minimum water connection flow pressure:
Cold water/steam condenser
Hot water (optional)
DI water (optional)
14.5
psi
5.8
psi
4.4
psi
100
kPa pressure
40
kPa pressure
30
kPa pressure
Recommended water connection flow pressure:
Cold water/hot water (optional)
DI water (optional)
Steam condenser
29
psi
29
psi
14.5
psi
≥
200
kPa pressure
≥
200
kPa pressure
≥
100
kPa pressure
Drain height
min.
11 3/4"
ft, max.
3' 3"
ft
Min.
0.3
m, max.
1.0
m
Drain hose length
max.
13' 1"
ft
Max.
4.0
m
Operation (according to IEC/EN 61010-1):
Ambient temperature
Relative humidity maximum
linear decrease to
Relative humidity minimum
40
°F to
104
°F
80 % for temperatures up to 88°F
50 % for temperatures up to 104°F
10%
5
°C to
40
°C
80% for temperatures up to 31°C
50% for temperatures up to 40°C
10%
Storage and transportation conditions:
Ambient temperature
Relative humidity
Air pressure
- 4
°F to
140
°F
10
% to
85
%
7.25
psi to
15.37
psi
- 20
°C to
60
°C
10
% to
85
%
500
hPa to
1060
hPa
Altitude above sea level (according to
IEC/EN 61010-1)
up to
4,921
ft*
Up to
2,000
m*
Type of protection (according to IEC 60529)
IP
21
Soiling level (according to IEC/EN 61010-1)
2
Overvoltage category (according to IEC 60664)
II
Sound emission values in dB (A),
sound pressure level LpA during cleaning and dry-
ing phases
<
70
Certifications
CAN/CSA-C22.2 No. 61010-1-04, CAN/CSA-C22.2 No. 61010-2-040, UL
Std. No. 61010-1 (2nd Edition), IEC 61010-2-040:2006
Manufacturer’s address
Miele & Cie. KG, Carl-Miele-Strasse 29,
33332 Gütersloh, Germany
* If installed above
4,921
ft (
2,000
m), the boiling point of the wash water will be lower. In this case, the disinfecting tempera-
ture and the holding time will need to be reset by
Miele
Customer Service.