Ophthalmology
Opthalmic instruments should be
cleaned and disinfected in a special
mobile injector unit.
The final rinse should be carried out
using demineralised water with a
conductivity of ~15
Z
S/cm
(Microsiemens per centimetre). In terms
of quality, the water used for the final
rinse must contain minimal amounts of
endotoxins and pyrogens.
The top level is fitted with various
connections for hollow instruments, e.g.
rinsing and suction hand pieces, and
cannulae.
Instruments are secured to the hose
connectors in the mobile injector unit
using the silicone holders and
connectors in the rack.
The lower level of the mobile injector
unit is designed to take E 441/1 inserts
or E 142 mesh trays for reprocessing
solid instruments.
,
Plastic fibre cover nets must not
be used in washer-disinfectors
which are used to reprocess hollow
ophthalmic instruments with narrow
lumen.
The mobile injector unit for
ophthalmic instruments is supplied
with its own operating instructions.
Areas of application
34