IFU-001 Rev D June 2014
FDA 510(k) Cleared
4. Install the gel pad on the electrode end of the CP Relief Wand®. (You can hold the CP-1000 with your
knees while applying the gel pads to the electrodes.) Visibly inspect that the electrode end of the device
is free of any contamination such as dust, dirt, or debris prior to application of the gel pad. Be sure that
the electrodes are centered in the gel pad windows.
5. With the Polarity control in the normal position and the Intensity control at zero set the Pulse Width
control to 150 uS. After you are familiar with using the 150 uS setting, you may wish to become familiar
with using the 230 uS setting for higher power.
6. With the Intensity control set to zero, turn the CP Relief Wand® power switch on and place the unit in
the required location. Press the CP Relief Wand® firmly against the skin and increase the Intensity to the
highest tolerable setting.
NOTE: You should always follow your
health professional’s instructions on proper use of the CP Relief
Wand®. The exact location of the pain is normally the best location to place the CP Relief Wand®. This
location can usually be found by pressing the finger in and around the area of the pain and locating the
most painful spot. The CP Relief Wand® is then pressed firmly over this spot. Gradually increase the
Intensity control to the maximum tolerable level. (If there is undesirable muscle vibration, the Intensity
can be reduced, the polarity can be reversed, the Wand can be rotated to a different angular position, or
the Wand can be moved to a different position.) Continue holding the CP Relief Wand® in this position
until some further reduction in pain is detected and again increase the intensity as high as is tolerable.
The device shall be used until pain subsides with a maximum treatment time of 30 minutes. Treatment
time may vary depending on the degree of pain, intensity setting,
and the patient’s tolerance to this
therapy. When using the CP Relief Wand®, the pain center may be larger than the width of the CP Relief
Wand®. In such case, to assure the entire pain center is treated, repeat the treatment process after
moving the CP Relief Wand® in ¾ inch steps on both sides of the original position. After completion of
treatment with the CP Relief Wand® the patient should feel a tingling which will go away in a short time.
Once an acceptable placement has been achieved, you may wish to record the site location in a
notebook or in your operation manual for future reference.
8. After use, replace the tape over the gel pad to keep it dry for the next use. For longer term storage,
place the gel pad back onto its original plastic film and inside the original sealed plastic bag.
IV. TECHNICAL SPECIFICATIONS
Peak Current
0-80 milliamps (500 ohms)
Peak Voltage
0-40 V (500 ohm)
Pulse Rate
40 Hz
Pulse Width
150 or 230 microseconds (selectable)
Waveform
Asymmetrical Bi-phase Square Pulse
Maximum Charge per pulse
20 Microcoulombs
Power Supply
9 Volt Alkaline Battery
Electrodes
Stainless Steel
FDA Cleared
FDA 510(k) #K133779
Safety Tests
Complies with IEC60601-1, IEC60601-1-2, and IEC60601-2-10
RoHS Compliance
Complies with RoHS Standards