4. Install the gel pad on the electrode end of the CP Relief Wand®. (You can hold the CP-1000 with your 
knees while applying the gel pads to the electrodes.) Visibly inspect that the electrode end of the device 
is free of any contamination such as dust, dirt, or debris prior to application of the gel pad.  Be sure that 
the electrodes are centered in the gel pad windows. 
 
5. With the Polarity control in the normal position and the Intensity control at zero set the Pulse Width 
control to 150 uS. After you are familiar with using the 150 uS setting, you may wish to become familiar 
with using the 230 uS setting for higher power.  
 
6. With the Intensity control set to zero, turn the CP Relief Wand® power switch on and place the unit in 
the required location. Press the CP Relief Wand® firmly against the skin and increase the Intensity to the 
highest tolerable setting. 
 
NOTE: You should always follow your 

health professional’s instructions on proper use of the CP Relief 

Wand®. The exact location of the pain is normally the best location to place the CP Relief Wand®. This 
location can usually be found by pressing the finger in and around the area of the pain and locating the 
most painful spot. The CP Relief Wand® is then pressed firmly over this spot. Gradually increase the 
Intensity control to the maximum tolerable level. (If there is undesirable muscle vibration, the Intensity 
can be reduced, the polarity can be reversed, the Wand can be rotated to a different angular position, or 
the Wand can be moved to a different position.) Continue holding the CP Relief Wand® in this position 
until some further reduction in pain is detected and again increase the intensity as high as is tolerable. 
The device shall be used until pain subsides with a maximum treatment time of 30 minutes.  Treatment 
time may vary depending on the degree of pain, intensity setting, 

and the patient’s tolerance to this 

therapy. When using the CP Relief Wand®, the pain center may be larger than the width of the CP Relief 
Wand®. In such case, to assure the entire pain center is treated, repeat the treatment process after 
moving the CP Relief Wand® in ¾ inch steps on both sides of the original position. After completion of 
treatment with the CP Relief Wand® the patient should feel a tingling which will go away in a short time. 
Once an acceptable placement has been achieved, you may wish to record the site location in a 
notebook or in your operation manual for future reference.  
 
8. After use, replace the tape over the gel pad to keep it dry for the next use. For longer term storage, 
place the gel pad back onto its original plastic film and inside the original sealed plastic bag.  

 
 
IV.        TECHNICAL SPECIFICATIONS 
 
 

Peak Current   

 

 

0-80 milliamps (500 ohms) 

Peak Voltage   

 

 

0-40 V (500 ohm) 

Pulse Rate 

 

 

 

40 Hz  

Pulse Width 

 

 

 

150 or 230 microseconds (selectable) 

Waveform 

 

 

 

Asymmetrical Bi-phase Square Pulse 

Maximum Charge per pulse   

20 Microcoulombs 

Power Supply  

 

 

9 Volt Alkaline Battery  

Electrodes 

 

 

 

Stainless Steel 

FDA Cleared   

 

 

FDA 510(k) #K133779 

Safety Tests   

 

 

Complies with IEC60601-1, IEC60601-1-2, and IEC60601-2-10 

RoHS Compliance 

 

 

Complies with RoHS Standards 

 
 
 
 
 

Summary of Contents for CP Relief WandCP 1000

Page 1: ...D June 2014 FDA 510 k Cleared CP Relief Wand Model CP 1000 Rx Only Mid America Medical Innovations LLC 2704 Industrial Drive Jefferson City MO 65109 USA 573 556 5569 844 226 1013 RoHS Compliant Opera...

Page 2: ...se Contraindications Precautions Cautions Warnings Adverse Reactions III Operation CP Relief Wand Components CP Relief Wand Controls CP Relief Wand Battery Information CP Relief Wand Electrodes Skin P...

Page 3: ...times almost everyone traveled long distances by foot and painful feet would have been a common problem Recovered ancient artifacts show that the Egyptians and Greeks knew of the pain reducing effects...

Page 4: ...atients 6 The long term effects of electrical stimulation are unknown 7 Since the effects of stimulation of the brain are unknown the unit s electrodes should not be applied anywhere on the head 8 The...

Page 5: ...on when the patient is in the bath or shower 7 Do not apply stimulation while the patient is sleeping or while the patient is driving operating machinery or during any activity in which electrical sti...

Page 6: ...th Switch The Pulse Width Switch selects either 150 microsecond pulse width low or 230 microsecond uS pulse width high Choose the 150 microsecond uS setting for lowest power and maximum comfort during...

Page 7: ...pads dry out you can moisten them with water for continued use After repeated usage the gel pads deliver less stimulation and should be replaced as needed Always use CP Relief Wand 2088 Gel Pads Skin...

Page 8: ...an be rotated to a different angular position or the Wand can be moved to a different position Continue holding the CP Relief Wand in this position until some further reduction in pain is detected and...

Page 9: ...Like any other stainless steel product after extended periods of use the stainless steel electrodes may tarnish due to the salts in your skin Stainless steel tarnish will not affect operation Should t...

Page 10: ...age resulting from failure to follow the operating instructions accidents abuse alteration or disassembly by unauthorized personnel When returning to the factory a Return Authorization Number must be...

Page 11: ...re its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class B The CP Relief Wand is suitable for use in all establishments...

Page 12: ...ble and mobile RF Communications equipment should be used no closer to any part of the device including cables than the recommended distance calculated from the equation applicable to the frequency of...

Page 13: ...ces are controlled The customer or user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitte...

Page 14: ...2014 FDA 510 k Cleared XI GLOSSARY Rx Sold by Prescription Only Caution Type BF Applied Part Manufactured by Mid America Medical Innovations LLC 2704 Industrial Dr Jefferson City MO 65109 USA uS Hz H...

Reviews: