IFU-001 Rev D June 2014 

 

FDA 510(k) Cleared 

 
CP Relief Wand® Controls 
 

The CP Relief Wand® controls are designed to be simple and easy to operate. Refer to the diagram 
below for the location of the controls. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
Intensity Control:

 The Intensity Control is a thumb dial located on the side for easy access during 

treatment. This control provides continuous adjustment and control of the amplitude of the pulsating 
electrical current during treatment. This dial is labeled 0-8 and corresponds to the output setting. Set the 
Intensity to zero before placing electrodes against the skin and reduce the Intensity to zero after 
removing the electrodes from the skin.   
 

On-Off Switch:

 This is the main power switch.   

 
Pulse Width Switch:

 The Pulse Width Switch selects either 150 microsecond pulse width (low) or 230 

microsecond (uS) pulse width (high). Choose the 150 microsecond (uS) setting for lowest power and 
maximum comfort during treatment.  
 

Polarity Switch:

 The Polarity Switch allows you to choose the polarity of the electrodes for best results. 

When using the CP Relief Wand® in certain locations of the body, uncomfortable muscle vibrations may 
occur during treatment. By changing the polarity (or rotating the electrode axis) uncomfortable muscle 
movements can often be minimized or eliminated.  

 
CP Relief Wand® Battery Information  
 

The battery in your CP Relief Wand® should last for up to 70 hours of operation and give you long 
lasting service. We recommend that the battery be replaced with a fresh battery at regular intervals to 
reduce the possibility of battery leakage. It is located in the CP-1000 handle and can be accessed by 

 

Summary of Contents for CP Relief WandCP 1000

Page 1: ...D June 2014 FDA 510 k Cleared CP Relief Wand Model CP 1000 Rx Only Mid America Medical Innovations LLC 2704 Industrial Drive Jefferson City MO 65109 USA 573 556 5569 844 226 1013 RoHS Compliant Opera...

Page 2: ...se Contraindications Precautions Cautions Warnings Adverse Reactions III Operation CP Relief Wand Components CP Relief Wand Controls CP Relief Wand Battery Information CP Relief Wand Electrodes Skin P...

Page 3: ...times almost everyone traveled long distances by foot and painful feet would have been a common problem Recovered ancient artifacts show that the Egyptians and Greeks knew of the pain reducing effects...

Page 4: ...atients 6 The long term effects of electrical stimulation are unknown 7 Since the effects of stimulation of the brain are unknown the unit s electrodes should not be applied anywhere on the head 8 The...

Page 5: ...on when the patient is in the bath or shower 7 Do not apply stimulation while the patient is sleeping or while the patient is driving operating machinery or during any activity in which electrical sti...

Page 6: ...th Switch The Pulse Width Switch selects either 150 microsecond pulse width low or 230 microsecond uS pulse width high Choose the 150 microsecond uS setting for lowest power and maximum comfort during...

Page 7: ...pads dry out you can moisten them with water for continued use After repeated usage the gel pads deliver less stimulation and should be replaced as needed Always use CP Relief Wand 2088 Gel Pads Skin...

Page 8: ...an be rotated to a different angular position or the Wand can be moved to a different position Continue holding the CP Relief Wand in this position until some further reduction in pain is detected and...

Page 9: ...Like any other stainless steel product after extended periods of use the stainless steel electrodes may tarnish due to the salts in your skin Stainless steel tarnish will not affect operation Should t...

Page 10: ...age resulting from failure to follow the operating instructions accidents abuse alteration or disassembly by unauthorized personnel When returning to the factory a Return Authorization Number must be...

Page 11: ...re its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class B The CP Relief Wand is suitable for use in all establishments...

Page 12: ...ble and mobile RF Communications equipment should be used no closer to any part of the device including cables than the recommended distance calculated from the equation applicable to the frequency of...

Page 13: ...ces are controlled The customer or user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitte...

Page 14: ...2014 FDA 510 k Cleared XI GLOSSARY Rx Sold by Prescription Only Caution Type BF Applied Part Manufactured by Mid America Medical Innovations LLC 2704 Industrial Dr Jefferson City MO 65109 USA uS Hz H...

Reviews: