3. Wait until it is indicated in the list of measurement instruments that the device is in setup mode. This is
indicated by a speech bubble with the "Setup" inscription.
4. Activate the display by briefly pressing (approx. 1 sec.) the solenoid switch.
5. During normal operation, the status indication in the centre of the display should now alternate
between the current gas measurement value and the current setpoint [mA] for the BLE mA Link .
Based on the currently displayed value, the type designation and serial number for the relevant module
are shown in the upper section of the display. If BLE paring of the data logger and BLE mA Link was
not possible, error "E15" is shown on the display.
Display of current H2S concentration Display of current setpoint
BLE pairing not possible
1
Type designation of the BLE module
3
Current value (incl. unit)
2
Serial number of the BLE module
6. The following steps are only necessary, if you simultaneously want to test the setpoint output.
7. Add a suitable measurement instrument (e.g. milliammeter) to the current loop and compare the
measured value with the one shown on the display of the data logger.
Schematic diagram: Checking the setpoint output
1
BLE mA Link
3
Actuator incl. power source for the current loop
(e.g. dosing pump controlled via analogue
signal)
2
Milliammeter
30
Rev. 01
Summary of Contents for BLE mA Link
Page 2: ......
Page 5: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 01 5 ...
Page 6: ...2 1 Declaration of conformity Appendix 6 Rev 01 ...
Page 8: ......
Page 16: ......
Page 19: ...Chapter 5 Functional principle Output current based on gas value Rev 01 19 ...
Page 20: ......
Page 22: ......
Page 34: ......
Page 36: ......
Page 38: ......
Page 40: ......
Page 42: ......
Page 44: ......