5. Installation, Assembly, Commissioning
5.1 Installation
Protect the instrument from:
•
sources of heat such as fan heaters or hot air outlets ;
•
extremely low temperatures;
•
direct sunlight;
•
excessive dust;
•
moisture or rain;
•
mechanical vibration or impacts;
•
equipment which contain magnets or generate magnetic fields.
5.2 Transportation Locking Pieces and Securements
Remove all locking pieces and adhesive tape used for securement during transportation.
•
Place the instrument on a rigid, vibration-free table or bench.
•
Align the instrument on the table. Lift up the instrument a little at the front.
•
Put the mains plug into the socket provided for this purpose.
•
See to correct voltage of your unit and of the mains.
Important! Allways use a 3-conductor, grounded wall outlet.
•
Connect the hand control panel to the socket provided for this purpose at the
rotary microtome (see figure).
•
Connect the optionally available foot switch.
•
Connect the optionally available flush-mounted illumination.
5.2.1 Rearview of the rotary microtome
Picture 3