MFE600C

 

Operation  Manual

 

V1.0   

 

Summary of Contents for MFE600C

Page 1: ...MFE600C Operation Manual V1 0...

Page 2: ...1 Report query 7 6 2 Parameters setting 7 6 3 Level 1 password menu 8 6 4 Level 2 menu 9 7 Installation 13 8 Modbus RTU communication 15 9 Responsibility 20 Our company reserves the right to modify th...

Page 3: ...n flowmeter can be constructed and installed without stopping the water It provides a new method for liquid flow measurement and installation It has been widely used in chemical water supply and drain...

Page 4: ...connected separately 3 1 Integrated wiring 1 2 3 4 5 6 7 8 9 10 11 12 13 14 1 15 T B RS232 T A IOUT POUT COM AH AL 3 3V S1 S0 1 S2 GND COM 24V IN 0V 1 2 Current Passive Active Pulse Passive Active RS2...

Page 5: ...are connected to Passive the current or pulse will output an active signal Active When lugs are connected to Active the current or pulse will output a passive signal 3 2 Separated wiring 220V 220V PE...

Page 6: ...5 Icom 24V Current output grounding wire 6 P 2 way flow pulse output frequency output 7 Pcom Pulse output grounding wire 8 ALMH Alarm output for Upper Limit of flow 9 ALML Alarm output for Lower Limit...

Page 7: ...up is shown in Table 4 Table 4 Display Description Reuse content Real time flow value and unit Null Real time flow rate and unit Click down key to switch display resistance value Alarm mark System ala...

Page 8: ...screen and 4 keys on the bottom side of the display screen Figure 4 Operation interface The meaning of the keys and operation methods are shown in Table 6 The composite key needs to press two keys at...

Page 9: ...ting and method 6 1 Report query In the main interface state Figure 4 press Function key to enter the report query interface Figure 5 and then select the corresponding function to view the data accord...

Page 10: ...00 4800 9600 14400 Check no check odd parity Check even parity check 3 Range setting m h L s L m L h Ukg s Ukg m Ukg h Usg s U sg m Usg h T s T m T h kg s kg m kg h m s m m 4 Damping time 0 5S 0 8S 1...

Page 11: ...1gal 0 01gal 0 001gal 1ig 0 1ig 0 01ig 0 001ig 13 Pulse output method Frequency output pulse output 14 Frequency output range 1 5000 15 LCD screen brightness 1 5 adjustable 6 4 Level 2 menu Users ent...

Page 12: ...mit alarm Yes No 4 Lower limit alarm value Full scale percentage setting 5 Current zero point correction adjustable 6 Current full scale correction adjustable 7 Excitation method 4 types 50 Hz 25 Hz 1...

Page 13: ...0 99S adjustable 14 Acquisition buffer depth 0 199 adjustable 15 Pressure sensor full scale 0 6MPa 1 0MPa 1 6MPa 16 Sensor coefficient value Factory set unchangeable 17 Additional coefficient Factory...

Page 14: ...23 Level 2 password modification adjustable 24 Date YY MM DD adjustable 25 Time HH MM SS adjustable 26 System running time System Run Time 1 Day 13 Hour 34 Min 43 Sec Hour Min Sec 1 1 5 7 4 9 Year Mon...

Page 15: ...owmeter in order to eliminate the influence of the gas at the top of the pipeline it is generally recommended to install the plug in electromagnetic flowmeter on a horizontal line or at a position of...

Page 16: ...ance of the first 10D and the rear 5D and the valve needs to be fully opened as shown in Figure 3 Figure 3 When the plug in electromagnetic flowmeter is installed at the back end of the T shaped tube...

Page 17: ...n instruction table register number is register address Composition of communication information Address code Function code Data segment CRC A message is sent and received continuously and the charact...

Page 18: ...then the low byte The high byte is the byte where the sign bit and exponent bit in the floating point number are located and the low byte is the byte where the mantissa is located Above D0 D1 D2 and D...

Page 19: ...D0 12H D1 ABH For floating point numbers the format is IEEE 754 single precision and the transmission order is the high byte first and then the low byte The high byte is the byte where the sign bit a...

Page 20: ...3 CRCL CRCH g Reverse total decimal places message definition FLOAT Host sends 01 03 00 0A 00 02 E4 09 Slave response 01 03 04 D0 D1 D2 D3 CRCL CRCH h Reverse total unit message definition SHORT List...

Page 21: ...ssage definition LONG Host sends 01 03 00 1A 00 02 E5 CC Slave response 01 03 04 D0 D1 D2 D3 CRCL CRCH t Reverse total integer high message definition LONG Host sends 01 03 00 1C 00 02 05 CD Slave res...

Page 22: ...0 0x00 0x00 FLOAT number represent 0 25 9 Responsibility Within one year from the delivery date we shall repair or replace the instrument with any quality fault caused by material parts or our manufac...

Page 23: ...www microsensorcorp com...