Microsemi Proprietary and Confidential. LX3301A/LX3302A Auto Calibration User Guide Revision 1.0
17
6. Check the box for
to get the MaxErr number that IPCE can achieve. Select
to
SIM_OUT_ERROR
Zoom
see error versus position. The following image shows an example simulation.
Figure 29 • Example Simulation
4.6
Save Calibrated Result into EEPROM
The following steps describe how to save the calibrated result into EEPROM.
1. When the auto calibration simulation result is acceptable, restore all test modes into normal mode.
2. Uncheck all boxes, except
on DIAGMK, as shown.
VIN POR Err
Figure 30 • Uncheck Boxes
3. For IO1, IO2, and IO3 Mode, select
, as shown.
0 =Normal
Figure 31 • Normal Mode
4. For DEBUG, select
, as shown.
00000=Disabled