Quadrus Verifier User’s Manual
3-41
Verification
Lighting
Definition:
A string of ASCII characters that represents the angle and configuration of
illumination used in the verification environment.
Important:
For reliable Direct Part Mark verification results, Microscan
recommends setting the
Lighting
parameter to
90
. This is because diffuse
perpendicular (or “on axis/bright field”) illumination—in which the symbol
plane is parallel to the plane of the Verifier’s sensor and the symbol is uniformly
illuminated at a 90° angle of incidence—is most effective for reading Direct
Part Marks.
Note:
Details about other lighting configurations and their corresponding
ASCII representations can be found in AIM Global’s “Direct Part Mark
(DPM) Quality Guideline”, available at
Serial Cmd:
<
K532,
minimum element size,maximum element size,wavelength,
lighting
>
Default:
45Q
Options:
Any ASCII string up to 15 characters
Summary of Contents for Quadrus Verifier
Page 1: ...Quadrus Verifier User s Manual P N 83 006702 Rev C ...
Page 30: ...1 18 Quadrus Verifier User s Manual Generate and Save Verification Report ...
Page 150: ...4 10 Quadrus Verifier User s Manual UII Parsing ...
Page 152: ...Quadrus Verifier User s Manual ...
Page 178: ...5 26 Quadrus Verifier User s Manual Aux Port System Data Status ...
Page 210: ...6 32 Quadrus Verifier User s Manual Setting Up the Verifier for EZ Trax ...
Page 302: ...8 62 Quadrus Verifier User s Manual Configuring EZ Trax Output ...
Page 318: ...9 16 Quadrus Verifier User s Manual New Master Pin ...
Page 326: ...10 8 Quadrus Verifier User s Manual Service Message ...
Page 362: ...12 14 Quadrus Verifier User s Manual Request All Database Settings ...
Page 390: ...14 22 Quadrus Verifier User s Manual Status Requests ...
Page 404: ...15 14 Quadrus Verifier User s Manual Ordered Output Filter ...
Page 444: ...A 30 Quadrus Verifier User s Manual Glossary of Terms ...