Quadrus Verifier User’s Manual
6-13
Read Cycle
Serial Trigger
Serial Trigger Character (Delimited)
Note:
Serial Data
or
Serial Data or Edge
triggering mode must be enabled for
Serial
Trigger Character
to take effect.
Start and Stop Trigger Characters (Non-Delimited)
Usage:
Allows the user to define the trigger character and delimiters that start and
stop the read cycle.
Definition:
A serial trigger
is considered an online host command and requires the
same command format as all host commands: that is, to be entered within
angle brackets delimiters
<
and
>
or in the case of non-delimited triggers,
define individual start and stop characters.
Usage:
Allows the user to define the trigger character that initiates the read cycle.
Definition:
A single ASCII host serial trigger character that initiates the read cycle.
A delimited trigger character is one that either starts or ends the read cycle
and is enclosed by delimiters such as
<
and
>
.
Serial Cmd:
<
K201,serial trigger character
>
Default:
Spacebar
(corresponds to
<SP>
displayed in the embedded menu)
Options:
Any single ASCII character, including control characters, except
NULL
(
00
),
an existing host command character, or an on-line protocol character. Con-
trol characters entered on the command line are displayed in the menu as
mnemonic characters.
Usage:
It is useful in applications where different characters are required to start
and end a read cycle.
Definition:
A non-delimited trigger character is one that either starts or ends the read
cycle and is NOT enclosed by delimiters such as
<
and
>
.
Both
Start
and
Stop
non-delimited characters can be defined and will
function according to the trigger event, as follows:
When defining
Start
and
Stop
trigger characters, the following rules apply:
• In
External Edge
the reader looks only for the start trigger character and
ignores any end trigger character that may be defined.
• In
External Level
, the start trigger character
begins the read cycle and
end trigger character
ends it. Note that even after a symbol has been
decoded and the symbol data transmitted, the reader remains in
External
Level
trigger read cycle until a
Stop
character is received.
• In
Serial Data and Edge
trigger mode, command, either a start trigger
character or a hardware trigger can start an edge trigger read cycle.
Summary of Contents for Quadrus Verifier
Page 1: ...Quadrus Verifier User s Manual P N 83 006702 Rev C ...
Page 30: ...1 18 Quadrus Verifier User s Manual Generate and Save Verification Report ...
Page 150: ...4 10 Quadrus Verifier User s Manual UII Parsing ...
Page 152: ...Quadrus Verifier User s Manual ...
Page 178: ...5 26 Quadrus Verifier User s Manual Aux Port System Data Status ...
Page 210: ...6 32 Quadrus Verifier User s Manual Setting Up the Verifier for EZ Trax ...
Page 302: ...8 62 Quadrus Verifier User s Manual Configuring EZ Trax Output ...
Page 318: ...9 16 Quadrus Verifier User s Manual New Master Pin ...
Page 326: ...10 8 Quadrus Verifier User s Manual Service Message ...