LVS-9580 Operating Instructions - English
LVS-9580 Operating Instructions - English
Page 12 of 17
5.
On the “Calibration” tab, make sure the blue line travels through the middle of the PASS
portion of the barcode (see Figure below).
6. C
lick the “Calibrate” button.
Successful calibration is indicated by a green “Calibration OK” message.
Failed calibration is indicated by a red
“Calibration Needed” message.
7. If calibration fails:
o
Re-scan the Master Grade barcode and follow the above steps to calibrate. It may take
two or three attempts before calibration is complete.
o
If calibration continues to fail, contact Microscan or your Microscan representative for
further instructions.
IMPORTANT:
The calibration score will hardly ever match exactly; this is normal
and acceptable as long as the scores are /- 3 percent.
8. When calibration is complete, click
the “Grading” tab to grade barcodes. See the next section
for steps on grading barcodes.
The blue line
must pass
through the
“PASS” portion
of the barcode
.