LVS-9580 Operating Instructions - English
LVS-9580 Operating Instructions - English
Page 11 of 17
3. Firmly grip the LVS-9580 handle. There is no need to press the trigger at this time (pressing
the trigger causes the LVS-9580 to capture a live image; however, the image is automatically
live while the system is in Calibration mode).
4. Place the LVS-9580 window over one of the Master Grade barcodes making sure the four
rubber feet surrounding the window rest firmly on a flat, stable surface. The rubber feet hold
the test card in place and prevent movement of the test card. See example of Master Grade
barcodes above. It is important to note that not all of the rubber feet will fit onto the calibration
card.
Not all of the
rubber feet will fit
onto the
calibration card.
Verifier Window