28
When Meter scans, the indicator will be automatically lights up first.
The white area in the rectangular light field is the sensing area. When
scanning, the scan port should be aligned with the barcode, and the
position, distance and angle between the barcode and scan port should
be adjusted according to the light field.
Location: The barcode should be placed in the white sensing area with the
barcode direction align with the long axis of the white sensing area.
Distance: Adjust the distance between the barcode and meter scan port
to make sure the barcode fits completely within the white sensing area.
Angle: The light field should be as perpendicular to the barcode plane as
possible, and the vertical deviation should have the highest recognition
rate within ±5°.
8.5 Install Codeship
Insert the test strip Codechip into
the chip slot (see Section 1.3 for
position details), then the specific
Codechip information will display
(Figure 3-5). Codechip number and
arrow side should be face up.
When the input Codechip number
of the test strip does not match the
Codechip information, the system
will deliver a warning message
indicating that the Codechip
information does not match.
3-5