6
netic radiation, as these may interfere with the proper operation of
the instrument.
5.3 Patient health Status
Current patient health status may cause inaccurate or unexpected test
results. It is important to take certain health factors into consideration
when interpreting the test results and deciding on a course of action for
your patients. Failure to do so may lead to an incorrect interpretation of
the test result.
5.4 Performing a Test
●
The qLabs
®
Vet should be placed on a level surface that is free
of vibration. Testing on an uneven or unstable surface may cause
inaccurate results. DO NOT hold the qLabs
®
Vet in your hands
during the testing.
●
Strictly follow the test procedure specified in the manual to perform
the test.
●
DO NOT move or touch the qLabs
®
Vet during testing.
●
After inserting the test strip, please confirm the consistency of the
code chip, failure to do may cause inaccurate test result.
●
The blood sample must be applied to the test strip immediately
after collection. Otherwise, the blood sample may begin clotting
and causing inaccurate results.
●
Do not add the sample until the system prompts you to start add-
ing sample.
●
The sample should be added in one continuous operation, do not
reapply additional sample.
●
Do not use the device for unintended purposes, failure to do so
may cause unexpected test results.
5.5 Collecting Sample
●
Use fresh venous whole blood and citrated whole blood. Do not
use plasma.
●
The drop of blood must be a minimum of 10 μL.
5.6 Warning Information on Built-in Lithium Battery
●
qLabs meter has a built-in lithium battery, which cannot be
removed by any user.