POSSIBLE SOURCES OF ERRORS – CAUSE AND REMOVAL
Problem
Cause Removal
Thick-thin-sections
−
Blunt knife/blade.
−
Knife angle, therefore,
unfavorable clearance angle.
−
Insufficient clamping on
specimen clamping and/or
knife carrier.
−
Move knife and/or insert a new
one.
−
Try clearance angle
adjustments, until an optimal
angle can be found.
−
Check all screw and clamping
connections
on
specimen
clamping and knife carrier.
Tighten them, if necessary.
Compressions
−
Blunt knife/blade.
−
Specimen too warm.
−
Unfavorable clearance angle.
−
Cutting speed too high.
−
Move knife and/or insert a new
one.
−
Cool specimen
−
Try clearance angle
adjustments until an optimal
angle can be found.
−
Turn handwheel slower.
"Chatter" on sections
−
Cutting speed too high.
−
Unfavourable clearance angle.
−
Insufficient clamping on
specimen clamping and/or
knife carrier.
−
Turn handwheel slower.
−
Try clearance angle
adjustments until an optimal
angle can be found.
−
Check all screw and clamping
connections
on
specimen
clamping and knife carrier.
Tighten them, if necessary.
No feed is carried out anymore,
and therefore, no sections are
produced.
−
Front end position has been
reached.
−
Coarse feed wheel is blocked.
−
Move specimen backwards by
turning the coarse feed wheel.
−
Remove blockade on coarse
feed wheel.
Tight handwheel
- partly
- during the entire movement
−
Debris and section waste
between microtome and base
plate.
−
Dirty link block.
−
Remove section waste and
clean microtome.
−
Contact a service technician.
Note:
In case of malfunctions and/or service work, please turn off the instrument and contact your
local dealer.