Rotary Microtome HM 325
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387821 - English
3-7
ACOUSTIC SIGNAL FOR FRONT
END POSITION OF THE
SPECIMEN MOVEMENT
•
About 1 mm before the front end position of
the specimen feed an acoustic signal sounds
during each rotation of the handwheel!
Warning:
When the signal sounds, a feed range
of no more than 1000 µm is available
to finish sectioning. To avoid damaging
mechanical parts, set back the specimen holder
by means of the coarse feed wheel before starting
the next sectioning series.
Note:
Sections should be cut in the back third
of the horizontal specimen movement
because the mechanical stability of the
guide cylinder is best in this area.