EN
23
«ABI» Mode
1)
Turn on the power
– Turn on the device by
switching the I/O switch at the back of the
device into the ON position.
2)
Switch to «ABI» Mode
– Slide the Mode Switch
to «
ABI»
Mode. The supine position reminder will
display.
3) The patient has to lie down in supine position.
4)
Apply the cuff*
– Properly it the arm cuff to the
upper arm and the ankle cuff to the leg of the
preferred measurement side of the body. The
preferred side is determined in
«SCREEN»
Mode.
5) Make sure the upper arm cuff is connected to the
left cuff socket and the ankle cuff is connected to
the right cuff socket.
DC 7.5V 2.0A
*
Additional visual instruction can be found on the cuff.
ABI ROUTINE SCREEN
IN
DE
X L
IN
E
8
L
E
F
ABI
L
R