EN
31
BT
Do not wash the cuff!
Do not iron the cuff!
Accuracy test
We recommend the WatchBP Home A BT device be
tested for accuracy every 2 years or after mechanical
impact (e.g. being dropped). Please contact Microlife to
arrange for an accuracy test.
Cuff cleaning and disinfecting
DO NOT wash the cuff. DO NOT iron the cuff cover.
Wipe the cuff with 70% ethyl or isopropyl alcohol. Do not
immerse hose. Allow to air-dry thoroughly before next
use.
Disposal
Batteries and electronic instruments must be
disposed of in accordance with the locally applicable
regulations, and not as domestic waste.
Device cleaning and disinfecting
Use a soft cloth with one of the following recommended
cleaning solutions to wipe the exterior of the device:
· Ethyl or isopropyl alcohol (70% solution)
· Hydrogen peroxide 7.5% solution
· Sodium hypochlorite solution (5.25-6.15% household bleach
diluted 1:500 provides >100 ppm available chlorine)