BP3KN1-3B
18
IB BP3KN1-3B V4920 EN
EN
Error
Description
Potential cause and remedy
«ERR 1»
Signal too Weak
The pulse signals on the cuff are too weak.
Reposition the cuff and repeat the
measurement.*
«ERR 2»
Error signal
During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«ERR 3»
No pressure in
the cuff
An adequate pressure cannot be generated in
the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is not
too loose. Replace the batteries if necessary.
Repeat the measurement.
«ERR 5»
Abnormal result
The measuring signals are inaccurate and no
result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«ERR 6»
MAM Mode
There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«ERR
21»
Heart Level
Indicator error
Possibly some problem occurs on the sensor
for Heart Level Indicator. Turn off and remove
the battery then repeat measurement. It still
displayed, please contact your local
Microlife-Service.
«HI»
Pulse or cuff
pressure too high
The pressure in the cuff is too high (over 299
mmHg) OR the pulse is too high (over 199
beats per minute). Relax for 5 minutes and
repeat the measurement.*
«LO»
Pulse too low
The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*