4
5. Safety, Care, Accuracy Test and Disposal
Safety and protection
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Never inflate beyond 300 mmHg.
Always deflate the cuff completely before storage.
Do not use this device if you think it is damaged or notice
anything unusual.
Never open this device.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING:
Do not wash the cuff in a washing machine or
dishwasher!
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Electronic devices must be disposed of in accordance with
the locally applicable regulations, not with domestic waste.
6. Guarantee
This device is covered by a
2 year guarantee
from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Tubes, pumping ball, cuff.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
7. Technical Specifications
Weight:
360 g
Dimensions:
175 x 70 x 103 mm
Storage conditions:
-20 - +70 °C
30 - 85 % relative maximum humidity
Operating conditions:
0 - 46 °C
Measurement range:
0 - 300 mmHg
Resolution:
2 mmHg
Static accuracy:
within ± 3 mmHg between 18 - 33 °C;
within ± 6 mmHg between 34 - 46 °C
Air leakage:
< ± 4 mmHg/min
Hysteresis error:
within 0 - 4 mmHg
Summary of Contents for BP AG1-10
Page 13: ...11 BP AG1 10 FR...
Page 19: ...17 BP AG1 10 IT...
Page 22: ...20 1 2 1 2 2 3 3 4 3 4 3 1 40 E 200 5 4 3 1 6 3 1 7 8 9 4 E 1 3 2 3 4 E 0 3 E 3...
Page 23: ...21 BP AG1 10 RU 5 300 2 Microlife 6 2 Microlife Microlife Microlife www microlife ru support...
Page 25: ...23 BP AG1 10 RU...
Page 35: ...33 BP AG1 10 HR...
Page 38: ...36 2 3 mmHg 1 2 1 1 2 1 2 2 3 4 3 4 3 1 40 200 5 4 1 6 1 7 8 9 4 1 3 4 2 3 0 3 3...
Page 41: ...39 BP AG1 10 BG...
Page 47: ...45 BP AG1 10 RO...
Page 49: ...47 BP AG1 10 AR ESH ESC JSH _...
Page 50: ...48...
Page 53: ...51 BP AG1 10 FA mmHg ESH ESC JSH mmHg _ 86082261...
Page 54: ...52 mmHg mmHg...
Page 55: ...53 BP AG1 10 FA mmHg 86082261 86082261 www microlife com support C mmHg C mmHg...
Page 56: ...54 mmHg mmHg EN ISO 81060 1 ANSI AAMI SP09 Directive 93 42 EEC...
Page 57: ...55 BP AG1 10 FA...