20
11. Technical data
Operating
10 - 40 °C / 50 - 104 °F
conditions:
15 - 95 % relative
maximum humidity
Storage
-20 - +55 °C / -4 - +131 °F
conditions:
15 - 95% relative
maximum humidity
Weight:
402 g (with batteries)
Size:
138 x 94.5 x 62.5 mm
Measuring method:
oscillometric, validated by
the Korotkoff method:
phase I systolic,
phase V diastolic
Measuring range:
20 - 280 mmHg – blood
pressure
40 - 200 beats
per minute – pulse
Cuff pressure
display range:
0 - 299 mmHg
Measuring
resolution:
1 mmHg
Static accuracy:
Pressure within ± 3 mmHg
Pulse rate accuracy:
± 5% of the readout value
Voltage source:
· 4 x 1.5 V alkaline
batteries, size AA
· Mains adapter DC 6V,
600 mA (optional)
Battery
ca. 920 measurements
life :
(with new batteries)
IP class:
IP 20
Reference to
EN 1060-1 /-3 /-4;
standards:
IEC 60601-1;
IEC 60601-1-2 (EMC);
IEC 60601-1-11
Average Monitor
: 5 years or
useful life:
10,000 mesurements;
Accessories: 2 years
This monitor complies with the requirements of the
Medical Device Directive 93/42/EEC.
Technical alterations reserved.