16
Error
Designa-
tion
Potential cause and
remedy
«ERR 5»
Irregular
measuring
result
The measuring signals
are inaccurate and no
result can therefore be
displayed. Read through
the checklist for per-
forming reliable measu-
rements and repeat the
measurement.*
«ERR 6»
MAM mode Too many errors oc-
curred during the mea-
surement in the MAM
mode so that no final
result can be determi-
ned. Read through the
checklist for performing
reliable measurements
and then repeat the
measurements.*
«HI»
Pulse or cuff
pressure too
high
The pressure in the cuff
is too high (over 299
mmHg) or your pulse
rate is too high (over 200
beats per minute). Relax
for 5 minutes and repeat
the measurement.*
Error
Designa-
tion
Potential cause and
remedy
«LO»
Pulse too
low
The pulse rate is too
low (less than 40 beats
per minute). Repeat the
measurement.*
* Please consult your doctor, if this or any other prob-
lem occurs repeatedly.
9 . Safety, care, accuracy test
and disposal
Safety and protection
·
Carefully follow these instructions. This
document gives important information on the
operation and safety of your monitor. Before
using the monitor, please read these instructions
carefully and retain them for future reference.
·
This device may only be used for the purposes
described in these instructions. The manufac-
turer will not accept any liability for damage
caused by improper use.
·
This blood pressure monitor contains sen-
sitive components and must be handled
carefully. Observe the storage and operating
conditions described in chapter 11 on
page 21.