for your supply voltage.
Ensure that neither the mains adapter nor the cable are damaged.
1.
Plug the adapter cable into the mains adapter socket (4) in the blood pressure monitor.
2.
Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.
12. Error Messages
If an error occurs during the measurement, the measurement is interrupted and an error message,
e.g. «
ERR 3
», is displayed.
Error
Description
Potential cause and remedy
«ERR 1»
Signal too
weak
The pulse signals on the cuff are too weak. Re-position the cuff and
repeat the measurement.*
«ERR 2»
(26)
Error signal
During the measurement, error signals were detected by the cuff,
caused for instance by movement or muscle tension. Repeat the
measurement, keeping your arm still.
«ERR 3»
(23)
No pressure in
the cuff
An adequate pressure cannot be generated in the cuff. A leak may
have occurred. Check that the cuff is correctly connected and is not
too loose. Replace the batteries if necessary. Repeat the
«ERR 5»
Abnormal result The measuring signals are inaccurate and no result can therefore be
displayed. Read through the checklist for performing reliable
measurements and then repeat the measurement.*
«ERR 6» AFIB/MAM
Mode
There were too many errors during the measurement in AFIB/MAM
mode, making it impossible to obtain a final result. Read through the
checklist for performing reliable measurements and then repeat the
«HI»
Pulse or cuff
pressure too
The pressure in the cuff is too high (over 300 mmHg) OR the pulse
is too high (over 200 beats per minute). Relax for 5 minutes and
«LO»
Pulse too low
The pulse is too low (less than 40 beats per minute). Repeat the
Please consult your doctor, if this or any other problem occurs repeatedly.
If you think the results are unusual, please read through the information in «Section 1.» carefully.
13. Safety, Care, Accuracy Test and Disposal
Safety and protection
This device may only be used for the purposes described in these instructions. The manufacturer
cannot be held liable for damage caused by incorrect application.
•
This device comprises sensitive components and must be treated with caution. Observe the
storage and operating conditions described in the «Technical Specifications» section.
•
Protect it from: