for your supply voltage.     

Ensure that neither the mains adapter nor the cable are damaged. 

1. 

Plug the adapter cable into the mains adapter socket (4) in the blood pressure monitor. 

2. 

Plug the adapter plug into the wall socket. 

When the mains adapter is connected, no battery current is consumed. 

 

12. Error Messages 

If an error occurs during the measurement, the measurement is interrupted and an error message, 

e.g. «

ERR 3

», is displayed. 

Error 

Description 

Potential cause and remedy 

«ERR 1» 

Signal too 

weak 

The pulse signals on the cuff are too weak. Re-position the cuff and 

repeat the measurement.* 

«ERR 2» 

(26) 

Error signal 

During the measurement, error signals were detected by the cuff, 

caused for instance by movement or muscle tension. Repeat the 

measurement, keeping your arm still. 

«ERR 3» 

(23) 

No pressure in 

the cuff 

An adequate pressure cannot be generated in the cuff. A leak may 

have occurred. Check that the cuff is correctly connected and is not 

too loose. Replace the batteries if necessary. Repeat the 

«ERR 5» 

Abnormal result  The measuring signals are inaccurate and no result can therefore be

displayed. Read through the checklist for performing reliable 

measurements and then repeat the measurement.* 

«ERR 6»      AFIB/MAM 

Mode 

There were too many errors during the measurement in AFIB/MAM 

mode, making it impossible to obtain a final result. Read through the

checklist for performing reliable measurements and then repeat the 

«HI» 

Pulse or cuff 

pressure too 

The pressure in the cuff is too high (over 300 mmHg) OR the pulse 

is too high (over 200 beats per minute). Relax for 5 minutes and 

«LO» 

Pulse too low 

The pulse is too low (less than 40 beats per minute). Repeat the 

Please consult your doctor, if this or any other problem occurs repeatedly.

 

If you think the results are unusual, please read through the information in «Section 1.» carefully. 

 

13. Safety, Care, Accuracy Test and Disposal 

 

Safety and protection 

This device may only be used for the purposes described in these instructions. The manufacturer 

cannot be held liable for damage caused by incorrect application. 

 

This device comprises sensitive components and must be treated with caution. Observe the 

storage and operating conditions described in the «Technical Specifications» section. 

 

Protect it from: 

Summary of Contents for AFIB BT

Page 1: ......

Page 2: ......

Page 3: ...Microlife AFIB BT BP A6 BT...

Page 4: ...Microlife AFIB BT BP A6 BT...

Page 5: ...tton 12 Forward Button 13 Lock Switch 14 USB Port Display 15 Date Time 16 Systolic Value 17 Diastolic Value 18 Pulse Rate 19 Battery Display 20 Traffic Light Display 21 Stored Value 22 Pulse Indicator...

Page 6: ...rithm of Microlife has been clinically investigated by several prominent clinical investigators and showed that the device detects patients with AFIB at a certainty of 97 100 1 2 Please read through t...

Page 7: ...de standard or AFIB MAM mode AFIB MAM mode highly recommended 4 Taking a Blood Pressure Measurement using this Device How not to store a reading 5 Appearance of the Atrial Fibrillation Indicator for e...

Page 8: ...s the day progresses You should therefore take your measurements in the same quiet conditions and when you feel relaxed Take at least two readings every time in the morning and in the evening and aver...

Page 9: ...rrhythmia or irregular heart beat It often causes no symptoms yet it significantly increases your risk of stroke You ll need a doctor to help you control the problem How does AFIB impact my family or...

Page 10: ...w you can enter the new values as described above Selecting the correct cuff Microlife offers different cuff sizes Select the cuff size to match the circumference of your upper arms measured by close...

Page 11: ...e which arm to measure in the future The arm with the higher blood pressure should be measured 4 Remove close fitting garments from the upper arm To avoid constriction shirt sleeves should not be roll...

Page 12: ...o cause for concern In such case it is recommended to measure again the next day However if the symbol appears on a regular basis e g several times a week with measurements taken daily we advise you t...

Page 13: ...crolife Blood Pressure APP microlife vitalis software The memory data can be transferred to the iPhone by connecting the monitor via a Bluetooth Installation and data transmission 1 Install the APP mi...

Page 14: ...of the device 3 Replace the batteries ensure correct polarity as shown by the symbols in the compartment The memory retains all values although date and time must be reset the year number therefore f...

Page 15: ...he measuring signals are inaccurate and no result can therefore be displayed Read through the checklist for performing reliable measurements and then repeat the measurement ERR 6 AFIB MAM Mode There w...

Page 16: ...device unsupervised some parts are small enough to be swallowed Device care Clean the device only with a soft dry cloth Cleaning the cuff Carefully remove spots on the cuff with a damp cloth and soap...

Page 17: ...quirements of the Medical Device Directive 93 42 EEC Technical alterations reserved 7KLV GHYLFH FRPSOLHV ZLWK 3DUW RI WKH UXOHV 2SHUDWLRQ LV VXEMHFW WR WKH IROORZLQJ WZR FRQGLWLRQV WKLV GHYLFH PD QRW...

Page 18: ...erent from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Changes or modifications not expressly approved by the party responsible for compli...

Reviews: