determined in this way is more reliable than one produced by a single measurement. AFIB 

detection is only activated in AFIB/ MAM mode. 

 

After pressing the START/STOP button (1), the selected AFIB/ MAM mode appears in the 

display as the MAM-symbol (25). 

 

The bottom, right hand section of the display shows a 1, 2 or 3 to indicate which of the 3 

measurements is currently being taken. 

 

There is a break of 15 seconds between the measurements (15 seconds are adequate according 

to «Blood Pressure Monitoring, 2001, 6:145-147» for oscillometric instruments). A count down 

indicates the remaining time. 

 

The individual results are not displayed. Your blood pressure will only be displayed after all 3 

measurements are taken. 

 

Do not remove the cuff between measurements. 

 

If one of the individual measurements was questionable, a fourth one is automatically taken. 

 

4. Taking a Blood Pressure Measurement using this device 

Checklist for taking a reliable measurement

 

1. Avoid activity, eating or smoking immediately before the measurement. 

2. Sit down for at least 5 minutes before the measurement and relax. 

3. Always measure on the same arm 

(normally left). It is recommended that doctors perform 

double arm measurements on a patients first visit in order to determine which arm to measure in 

the future. The arm with the higher blood pressure should be measured. 

4

Remove close-fitting garments from the upper arm. To avoid constriction, shirt sleeves should 

not be rolled up – they do not interfere with the cuff if they are laid flat. 

5. Always ensure that the correct cuff size is used (marking on the cuff). 

 

Fit the cuff closely, but not too tight. 

 

Make sure that the cuff is positioned 2 cm above the elbow. 

 

The 

artery mark 

located on the cuff (ca. 3 cm long bar) must lie over the artery which runs           

down the inner side of the arm. 

 

Support your arm so it is relaxed. 

 

Ensure that the cuff is at the same height as your heart. 

6. Slide the lock switch (13) down to the «unlock» position. Press the START/STOP  button (1) to 

start  measuring. 

7. The cuff will now pump up automatically. Relax, do not move and do not tense your arm 

muscles until the measurement result is displayed. Breathe normally and do not talk. 

8. When the correct pressure is reached, the pumping stops and the pressure falls gradually. If the 

required pressure was not reached, the device will automatically pump some more air into the cuff. 

9. During the measurement, the pulse indicator (22) flashes in the display. 

10 The result, comprising the systolic (16) and the diastolic (17) blood pressure and the pulse rate (18) 

is displayed. Note also the explanations on further display symbols in this booklet.   

11. When the device has finished measuring, remove the cuff. 

Summary of Contents for AFIB BT

Page 1: ......

Page 2: ......

Page 3: ...Microlife AFIB BT BP A6 BT...

Page 4: ...Microlife AFIB BT BP A6 BT...

Page 5: ...tton 12 Forward Button 13 Lock Switch 14 USB Port Display 15 Date Time 16 Systolic Value 17 Diastolic Value 18 Pulse Rate 19 Battery Display 20 Traffic Light Display 21 Stored Value 22 Pulse Indicator...

Page 6: ...rithm of Microlife has been clinically investigated by several prominent clinical investigators and showed that the device detects patients with AFIB at a certainty of 97 100 1 2 Please read through t...

Page 7: ...de standard or AFIB MAM mode AFIB MAM mode highly recommended 4 Taking a Blood Pressure Measurement using this Device How not to store a reading 5 Appearance of the Atrial Fibrillation Indicator for e...

Page 8: ...s the day progresses You should therefore take your measurements in the same quiet conditions and when you feel relaxed Take at least two readings every time in the morning and in the evening and aver...

Page 9: ...rrhythmia or irregular heart beat It often causes no symptoms yet it significantly increases your risk of stroke You ll need a doctor to help you control the problem How does AFIB impact my family or...

Page 10: ...w you can enter the new values as described above Selecting the correct cuff Microlife offers different cuff sizes Select the cuff size to match the circumference of your upper arms measured by close...

Page 11: ...e which arm to measure in the future The arm with the higher blood pressure should be measured 4 Remove close fitting garments from the upper arm To avoid constriction shirt sleeves should not be roll...

Page 12: ...o cause for concern In such case it is recommended to measure again the next day However if the symbol appears on a regular basis e g several times a week with measurements taken daily we advise you t...

Page 13: ...crolife Blood Pressure APP microlife vitalis software The memory data can be transferred to the iPhone by connecting the monitor via a Bluetooth Installation and data transmission 1 Install the APP mi...

Page 14: ...of the device 3 Replace the batteries ensure correct polarity as shown by the symbols in the compartment The memory retains all values although date and time must be reset the year number therefore f...

Page 15: ...he measuring signals are inaccurate and no result can therefore be displayed Read through the checklist for performing reliable measurements and then repeat the measurement ERR 6 AFIB MAM Mode There w...

Page 16: ...device unsupervised some parts are small enough to be swallowed Device care Clean the device only with a soft dry cloth Cleaning the cuff Carefully remove spots on the cuff with a damp cloth and soap...

Page 17: ...quirements of the Medical Device Directive 93 42 EEC Technical alterations reserved 7KLV GHYLFH FRPSOLHV ZLWK 3DUW RI WKH UXOHV 2SHUDWLRQ LV VXEMHFW WR WKH IROORZLQJ WZR FRQGLWLRQV WKLV GHYLFH PD QRW...

Page 18: ...erent from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Changes or modifications not expressly approved by the party responsible for compli...

Reviews: