background image

4

Only use Microlife cuffs.

Contact your local Microlife Service if the enclosed cuff 

6

 does 

not fit.

Connect the cuff to the device by inserting the cuff connector 

7

 

into the cuff socket 

3

 as far as it will go.

Spare part cuffs are delivered with different cuff connectors 

7

If you exchange/replace the cuff for this device please ensure 

that is the same connector as the one supplied with the original 

cuff.

Selecting standard or MAM mode

This device enables you to select either standard (standard single 

measurement) or MAM mode (automatic triple measurement). To 

select standard mode, slide the MAM switch 

8

 on the side of the 

device downwards to position «1» and to select MAM mode, slide 

this switch upwards to position «3».

MAM mode

In MAM mode, 3 measurements are automatically taken in succession 

and the result is then automatically analysed and displayed. Because 

the blood pressure constantly fluctuates, a result determined in this 

way is more reliable than one produced by a single measurement. 

When you select the 3 measurements, the MAM-symbol 

AS

 

appears in the display.

The bottom, right hand section of the display shows a 1, 2 or 3 to 

indicate which of the 3 measurements is currently being taken.

There is a break of 15 seconds between the measurements. 

A count down indicates the remaining time.

The individual results are not displayed. Your blood pressure 

will only be displayed after all 3 measurements are taken.

Do not remove the cuff between measurements.

If one of the individual measurements was questionable, a 

fourth one is automatically taken.

Microlife AFIB is also activated in standard mode. If the 

device detects AF in standard mode, the cuff pressure will 

stop at a level of approx. 60 mmHg to receive more data. 

The hourglass 

BK

 will appear in the display.

4. Taking a Blood Pressure Measurement

Checklist for taking a reliable measurement

1. Avoid activity, eating or smoking immediately before the 

measurement.

2. Sit down on a back-supported chair and relax for 5 minutes. 

Keep the feet flat on the floor and do not cross your legs.

3.

Always measure on the same arm

 (normally left). It is recom-

mended that doctors perform double arm measurements on a 

patients first visit in order to determine which arm to measure in 

the future. The arm with the higher blood pressure should be 

measured.

4. Remove close-fitting garments from the upper arm. To avoid 

constriction, shirt sleeves should not be rolled up - they do not 

interfere with the cuff if they are laid flat.

5. Always ensure that the correct cuff size is used (marking on the 

cuff).

Fit the cuff closely, but not too tight.

Make sure that the cuff is positioned 2 cm above the elbow.

The 

artery mark

 on the cuff (ca. 3 cm long bar) must lie over 

the artery which runs down the inner side of the arm.

Support your arm so it is relaxed.

Ensure that the cuff is at the same height as your heart.

6. Slide the lock switch 

9

 down to the «unlock» position. Press 

the START/STOP button 

1

 to start measuring.

7. The cuff will now pump up automatically. Relax, do not move 

and do not tense your arm muscles until the measurement 

result is displayed. Breathe normally and do not talk.

8. When the correct pressure is reached, the pumping stops and the 

pressure falls gradually. If the required pressure was not reached, 

the device will automatically pump some more air into the cuff.

9. During the measurement, the pulse indicator 

AP

 flashes in the 

display.

10.The result, comprising the systolic 

AT

 and the diastolic 

AK

 

blood pressure and the pulse rate 

AL

 is displayed. Note also the 

explanations on further display symbols in this booklet.

11.When the device has finished measuring, remove the cuff.

12.Switch off the device. (The monitor does switch off automati-

cally after approx. 1 min.).

You can stop the measurement at any time by pressing the 

START/STOP button (e.g. if you feel uneasy or an 

unpleasant pressure sensation). 

If the systolic blood pressure is known to be very high

it can be an advantage to set the pressure individually. 

Press the START/STOP button after the monitor has been 

pumped up to a level of approx. 30 mmHg (shown on the 

display). Keep the button pressed until the pressure is 

about 40 mmHg above the expected systolic value – then 

release the button. 

Summary of Contents for AFIB ADVANCED Easy

Page 1: ...sy IB BP A6 Advanced Easy IT V4 4818 BP A6 Advanced Easy Colpharmas r l Parma Italy www colpharma com Distributoreperl Italia MicrolifeAG Espenstrasse139 9443Widnau Switzerland www microlife com EN 1...

Page 2: ...ced Easy BP A6 Easy Tagliando di garanzia Name of Purchaser Nome del rivenditore Nom de l acheteur Name des K ufers Serial Number Numero di serie Num ro de s rie Serien Nr Date of Purchase Data d acqu...

Page 3: ...algorithm is recommended for people of 65 years and older The AFIB algorithm indicates that atrial fibrillation may be present For this reason it is recommended that you visit your doctor when the dev...

Page 4: ...ssure of the blood flowing in the arteries generated by the pumping of the heart Two values the systolic upper value and the diastolic lower value are always measured The device indicates the pulse ra...

Page 5: ...DIA higher than 99 as well as those with diabetes coronary heart failure or for those who have previously had a stroke In young people or in pregnancy AF screening is not recom mended as it could gene...

Page 6: ...t flat on the floor and do not cross your legs 3 Always measure on the same arm normally left It is recom mended that doctors perform double arm measurements on a patients first visit in order to dete...

Page 7: ...batteries 1 Open the battery compartment 5 on the bottom of the device 2 Replace the batteries ensure correct polarity as shown by the symbols in the compartment Which batteries and which procedure Us...

Page 8: ...ce close to strong electromagnetic fields such as mobile telephones or radio installations Keep a minimum distance of 3 3 m from such devices when using this device Do not use this device if you think...

Page 9: ...e guarantee is valid only on presentation of the guar antee card completed by the dealer see back confirming date of purchase or the receipt Batteries and parts that become worn with use are not inclu...

Page 10: ...IB indica una possibile presenza di Fibrillazione Atriale Per questo motivo raccomandiamo di farsi visitare dal medico se l apparecchio segnala con regolarit la Fibrillazione Atriale La tecnologia AFI...

Page 11: ...nelle arterie generata dal pompaggio del cuore Si misurano sempre due valori quello sistolico massima e quello diasto lico minima Il dispositivo indica anche la frequenza cardiaca il numero di battiti...

Page 12: ...precedente ICTUS Nelle persone giovani o in gravidanza lo screening della Fibrilla zione Atriale non raccomandato in quanto potrebbe generare risultati falsamente positivi e provocare inutile ansia In...

Page 13: ...zione evitare di fare attivit fisica mangiare o fumare 2 Stare seduti su una sedia con schienale e rilassarsi per 5 minuti prima della misurazione Mantenere i piedi ben appoggiati a terra e non accava...

Page 14: ...lizzazione batteria parzialmente carica Anche se il dispositivo effettuer le misurazioni in modo affidabile necessario sostituire le batterie Batterie esaurite sostituzione Quando le batterie sono esa...

Page 15: ...to descritte nel capitolo Specifiche tecniche Proteggere il dispositivo da acqua e umidit temperature estreme urti e cadute contaminazione e polvere luce solare diretta caldo e freddo I bracciali sono...

Page 16: ...uzione Smaltimento Le batterie e gli strumenti elettronici devono essere smaltiti in conformit alle disposizioni locali e non come i rifiuti domestici 12 Garanzia Questo dispositivo coperto da una gar...

Page 17: ...15 BP A6 Advanced Easy IT...

Page 18: ...emarquez aucun sympt me caract ristique de cette affection Le d pistage de la FA en g n ral et aussi avec l algorithme Microlife AFIB est recommand pour les personnes de 65 ans et plus L algorithme AF...

Page 19: ...adaptateur secteur 10 Messages d erreurs 11 S curit entretien test de pr cision et limination de l quipement S curit et protection Entretien de l appareil Nettoyage du brassard Test de pr cision limin...

Page 20: ...s rapides et irr guliers C est ce qu on appelle la fibril lation La fibrillation atriale est la forme la plus courante d arythmie cardiaque Vous pouvez vivre normalement avec cette arythmie mais elle...

Page 21: ...chent pas Votre tension n appara tra sur l cran qu apr s la r alisation des 3 mesures N enlevez pas le brassard entre les mesures Si l appareil juge l une des mesures de la s rie non plausible il en e...

Page 22: ...u rectangle la valeur d affichage est soit l int rieur de l optimum vert lev e jaune trop haute orange ou dangereusement haute gamme rouge Cette classification correspond 4 plages d finies par les dir...

Page 23: ...lez consulter imm diatement votre m decin si cela ou tout autre probl me se produit plusieurs reprises Si vous obtenez des r sultats que vous jugez inhabituels veuillez lire attentivement les indicati...

Page 24: ...tez pas uniquement sur le r sultat de la mesure consid rez toujours d autres sympt mes pouvant survenir et le ressenti du patient Il est conseill d appeler un m decin ou une ambulance si n cessaire Ne...

Page 25: ...me la m thode Korotkoff phase I systolique phase V diastolique Etendue de mesure 20 280 mmHg tension 30 200 battements par minute pouls Plage de pression affich e du brassard 0 299 mmHg R solution 1 m...

Page 26: ...ls auch mit dem Microlife AFIB Algorithmus ab einem Alter von 65 Jahren oder lter empfohlen Der AFIB Algo rithmus zeigt an dass Vorhofflimmern vorhanden sein kann Aus diesem Grunde empfehlen wir Ihnen...

Page 27: ...Wichtige Informationen zum Blutdruck und zur Selbstmessung Blutdruck ist der Druck des in den Blutgef ssen fliessenden Blutes verursacht durch das Pumpen des Herzens Es werden immer zwei Werte gemess...

Page 28: ...s mit dem Alter zunimmt VHF Screening wird bei Personen ab dem 50 Lebens jahr empfohlen die hohen Blutdruck z B SYS ber 159 oder DIA h her als 99 Diabetes koronare Herzinsuffizienz oder bereits zuvor...

Page 29: ...n Sie die Manschette zwischen den Messungen angelegt Falls eine Einzelmessung fehlerhaft war wird automatisch eine vierte Messung durchgef hrt AFIB ist auch im Normal Modus aktiviert Wenn das Ger t Vo...

Page 30: ...oder gef hrlich hoch rot Die Klassifizie rung orientiert sich an den 4 Bereichen in der Tabelle der internatio nalen Richtlinien ESH AHA JSH wie in Kapitel 1 beschrieben 7 Messwertspeicher Dieses Ger...

Page 31: ...ung z B ERR 3 angezeigt Sprechen Sie umgehend mit Ihrem Arzt wenn diese oder andere Probleme wiederholt auftreten sollten Wenn Ihnen die Ergebnisse ungew hnlich erscheinen beachten Sie bitte sorgf lti...

Page 32: ...lem dann nicht wenn das Ergebnis nicht dem Befinden des Patienten entspricht Verlassen Sie sich nicht ausschliesslich auf das Messergebnis Alle potenziell auftretenden Symptome und die Schilderung des...

Page 33: ...risch validiert nach Korotkoff Methode Phase I systolisch Phase V dias tolisch Messbereich 20 280 mmHg Blutdruck 40 200 Schl ge pro Minute Puls Displaybereich Manschetten druck 0 299 mmHg Messaufl sun...

Reviews: