background image

 

The number of device feeds and the data buffer 

3.1.7  Retry count: 

The retry count specifies how many times the Web-Logger should retry obtaining the data from the 
device after a timeout has occurred. The Web-Logger will re-request the data from the device 
consecutively until a successful data request has occurred or until the number of re-tries are 
exceeded. 
 
Click "Save" to add the device. 
 
Repeat these steps for each device feed you want to add. 
 

3.2 

Avoiding data request delays: 

A data request delay occurs when the time it takes to request data from all the listed device feeds 
takes longer to acquire than the set request interval on a single device. 

For example, assuming the following configuration has been captured for each device feed: 

 
Number of device feeds 

: 10 

Request Interval 

 

: 5 seconds 

Timeout value 

 

: 500ms (0.5 seconds) 

Retry count   

 

: 3 

 
In order to obtain the data for all the listed device feeds, assuming each device responds within 
300ms, it would take in total: 
 

10 device feeds x 300ms = 3000ms (3 seconds) 

 
Assuming one of the devices that has 3 device feeds set, is not connected and causes a timeout. The 
retry count of 3 has been set. The time it takes to query this specific device will then be: 
 

3 retries x 500ms = 1500ms x 3 feeds = 4500ms (4.5 seconds) 

 
The total time to obtain data from all the devices would then be as follows: 
 

9 device feeds x 100ms  

= 900ms 

1 time out device     

= 4500ms 

 

 

 

 

= 5400ms (5.4 seconds) 

 
The total time exceeds the 5 second interval selected initially. This will cause the Web-logger to take 
longer than expected to process the list of device feeds and will result in the data not being sampled 
every 5 seconds. 
In order to correct this situation: 

1.  Attend to the non-communication device 
2.  Lower the timeout 
3.  Lower the retry count 
4.  Extend the request interval 

 

4. 

The number of device feeds and the data buffer 

The number of device feeds as well as their request interval and the upload interval of the Web-
Logger will determine how quickly the Web-Logger buffer will be filled up 
 
As per the above example, if you have 10 device feeds set to request every 5 seconds, a total of 10 
data samples will be buffered every 5 seconds. If the upload interval is set to every 1 minute, a total 
of 120 data samples will be stored in the buffer: 
 
 

Total requests per device feed in 1 minute: 

Summary of Contents for Web-Logger

Page 1: ...WEB LOGGER User Manual Manual Version WL 2016 1...

Page 2: ...Address 7 2 1 3 Friendly Name 7 2 1 4 Port to use 7 2 1 5 Data Buffer 7 2 1 6 Upload Size 7 2 1 7 Upload Type 8 3 Registering Devices 9 3 1 Adding a new device feed 9 3 1 1 Select Weblogger 10 3 1 2...

Page 3: ...on the router 4 Microcare Battery Monitor or at least one Microcare device with a MC Coms converter Connecting up 1 1 Standalone no Microcare Battery Monitor wiring diagram In this configuration a sin...

Page 4: ...1 2 In conjunction with a Microcare Battery Monitor wiring diagram In this configuration the Web Logger Battery Monitor and RS232 485 converters all connect in series creating the Microcare communicat...

Page 5: ...positive side to your battery positive and the negative side to your battery negative max supply voltage to not exceed 70v DC Plug the Ethernet cable into the silver Ethernet port of the Web Logger b...

Page 6: ...za Log in using your credentials or select the register if you do not already have an account Enter the required details and click the Register Once you have created and account and logged onto the w...

Page 7: ...records from the buffer will be uploaded at the same time For example If the data buffer is full 1 000 100 records at a time will be uploaded in order to clear the buffer over 10 upload intervals 100...

Page 8: ...escribed below 2 1 7 Upload Type Currently the All Data option is the only supported option Click save to save the changes If the WebLogger has a working internet connection you will notice the last c...

Page 9: ...tell it what data to acquire from each device On the website menu bar select the Devices option Here a list of Microcare monitored device feeds will be shown If you have more than one Web Logger the...

Page 10: ...e network address is the unique address of the device you are capturing Each device on the Microcare communications network must have a unique address The default for a Web Logger is 1 MPPT 5 Inverter...

Page 11: ...3 device feeds set is not connected and causes a timeout The retry count of 3 has been set The time it takes to query this specific device will then be 3 retries x 500ms 1500ms x 3 feeds 4500ms 4 5 s...

Page 12: ...o the Inputs menu option Once the Web Logger has determined the device feeds required it will commence with requesting the data from the specific devices and will then upload the data to the Powermoni...

Page 13: ...ole remedies Destrier Electronics shall not be liable for damages whether direct incidental special or consequential even caused by negligence or fault Destrier Electronics owns all parts removed from...

Page 14: ...ct Number E mail Address Details of Product Owner Name Surname Address City Province Contact Number E mail Address Date Installed Microcare 1st Floor Neave Industrial Park Korsten Port Elizabeth P O B...

Reviews: