WHAT TO DO IF THE
RED LED
COMES ON
•
If the red LED flashes very briefly, less than one second, that
just means the system is changing modes within a protocol.
This is nothing to worry about.
•
If the red LED comes on for a second or two, that means the
system has detected a potential OVER TEMPERATURE
condition. It may correct itself, or you may need to reduce
the frequency or intensity of your protocol.
•
If the red LED comes on and stays on, that means a CRITICAL
OVER TEMPERATURE condition has been detected. Your
system is definitely running too hot and it has decided to shut
down. You may see a warning screen. You should turn the
power OFF. Check the coils: be sure they are plugged in all
the way. A slightly pulled out coil may cause a short.
•
The system may automatically adjust your intensity (power
level) down to a safe level, usually level 9.
ADVANCED SYSTEM DIAGNOSTICS
COPYRIGHT © 2018 Micro-Pulse LLC
Channel 1 over temp
Channel 2 over temp
Channel 3 over temp
Channel 4 over temp