Once RUNNING, press the BLUE (DOWN) button to view the
status of the coils. Four possible status indicators are shown:
“OK” “Short” “No Coil” “Over Temp”
•
You may have to PRESS and HOLD the BLUE button for one
second or longer for this screen to come up.
•
Here are the possible indications:
•
OK
- this is what you should see most of the time. It means
everything is OK.
•
Short
- this usually means your coil is not pushed in far enough.
Push it in.
•
No Coil
- this usually means you do not have a coil plugged in, or a
wire is broken.
•
Over Temp
–
This means the system thinks the internal
temperature is getting too high
•
If the temperature is persistently too high, the C5 will warn you
with a new screen, then it will flash the RED LED. Eventually the
system will shut down the output power in an attempt to prevent
damage. If you see this, try reducing the frequency or intensity.
ADVANCED SYSTEM DIAGNOSTICS
COPYRIGHT © 2018 Micro-Pulse LLC
Channel 1 OK
Channel 2 Short
Channel 3 No Coil
Channel 4 over temp
PRESS
and
HOLD
DO NOT WORRY ABOUT THESE SCREENS UNLESS THERE HAS BEEN A MAJOR MALFUNCTION. Call Micro-Pulse for troubleshooting