
Note
Follow the three coil placement guidelines to make sure you are receiving the full benefit from
the magnetic fields.
Visualizing the magnetic field can be helpful when considering how to place the coils. When the coils are placed in an Opposite-Side
configuration, they will generate a magnetic field that fills the space between them. The closer together you place the coils, the
stronger the field will be, and the more stimulation you will be able to focus onto the injured tissue. When the coils are placed in a
Side-by-Side configuration, they will generate a magnetic field that is wider across the area. The closer the two adjacent coils are,
the stronger the field will be. It is therefore recommended that the coils touch each other in the Side-by-Side configuration.
As stated in the guidelines, place the bumpy side of each coil away from the skin. This applies to both the Opposite-Side and Side-
by-Side configurations. This ensures the magnetic fields are lined up properly. If the bumpy sides are not out, then the magnetic
fields may oppose and cancel one another, and will render the system ineffective.
It is okay if the wound is larger than can be completely covered by the coils. The stimulation has a beneficial effect on tissues in
the general area of stimulation, so coil placements do not need to be precise.
It may also be helpful to move and reposition the coils occasionally, especially if the exact site of the injury is not clear.
Repositioning will change the orientation of the magnetic field, which is often beneficial in the recovery process. For example, on
the first day or two you might place the coils on a limb on the lateral (outside) and medial (inside) surfaces. Then for the following
day or two you might choose to reposition the coils on the dorsal (back) and ventral (front) surfaces around the injury. You might
later decide to place them obliquely (at an angle).
To the extent possible, it is best to try several options to determine what seems to work best. The preferred placement of the coils
might ultimately be driven mainly by practical concerns such as ease of securing or bandaging the coils or placement of the pulse
generator.
You should not use the device on or around the brain, ears, and eyes, unless you are under specific instructions from a
medical professional.
WARNING
Do not use the Alleva-Wave on or around the brain, ears, and eyes unless under specific instructions
from a medical professional.
The following major section of this manual shows the coils in various positions and configurations as examples.
Secure the Coils and the Pulse Generator
The two coils are normally placed directly against the skin and secured in
place using adhesive medical tape or medical bandages. Loose clothing can be
worn over the coils, if their location permits.
Alternatively, the coils can be located outside the clothing and fixed in place
using tape or elastic bands. In this case the clothing and coils at the location
should be held close to the skin for maximum magnetic field penetration into the
injured area.
The magnetic fields will penetrate clothing, tape, bandages, leather and other
non-metallic materials with no problems. However, the magnetic fields can be
distorted or weakened by the presence of metallic objects. These can include
watches, metal zippers, belt buckles, coins, cell phones, and other electronic
devices. You should keep metallic objects at least 8 inches (200 mm) away from
the coils.
Note
Keep metallic objects at least 8 inches (200 mm) away from the coils to ensure the magnetic fields
are properly focused.
The pulse generator can similarly be secured by adhesive medical tape or medical bandages. It can also be placed in a pocket if one
is available. Some users fashion a special simple garment with extra pockets or pouches to hold the coils and pulse generator in
www.allevawave.com
7 of 12
P2 User Instruction Manual