Stimulation and rest periods
Periods of rest are just as important as periods of stimulation to allow tissues to recover. The Alleva-Wave gives you two options
to manage stimulation and rest periods. One option is for the Alleva-Wave to manage them for you. By moving the cycle profile
switch to position 1 or 2, the P2 runs on an automatic and continuous 41 minute cycle, which includes a 30 minute stimulation
period (Stimulation Mode) followed by an 11 minute rest period (Rest Mode). The 41 minute cycle repeats automatically; you do
not have to do anything. This option is useful if you want to wear the device for many hours at a time without managing your
stimulation and rest times.
The other option is for you to manage the periods of stimulation and rest. By moving the cycle profile switch to position 3, the
Alleva-Wave runs continuously in Stimulation Mode with no rest periods. This option can be used when you want to have a brief
and continuous stimulation period for perhaps an hour or two.
Two levels of magnetic field intensity
The Alleva-Wave can generate either standard intensity magnetic fields for mild tissue stimulation, or higher intensity fields for
vigorous tissue stimulation. You can move the cycle profile switch to position 1 to generate standard intensity fields, or you can
select position 2 or 3 to generate higher intensity fields.
When first using the Alleva-Wave it is recommended that you set the cycle profile to 1 (standard intensity with rest periods) on the
first day or two as you become familiar with the device. Thereafter you can select cycle profile 2 (higher intensity with rest periods)
or 3 (higher intensity with no rest periods).
Higher intensity fields are recommended when the injured tissue is far from the coils. It is also recommended after you feel
comfortable with low intensity fields and you feel that more stimulation intensity would be beneficial for your tissue.
Cycle profile switch
The Alleva-Wave P2 operates on a 41 minute cycle that repeats continuously and automatically. Using the cycle profile switch,
you can select from among four predefined cycle profiles.
Switch
Magnetic
Mode for
Mode for
position
field intensity
first 30 minutes
next 11 minutes
0
OFF
OFF
OFF
1
Standard
Stimulation (flashing green)
Rest (flashing yellow)
2
Higher
Stimulation (flashing green)
Rest (flashing yellow)
3
Higher
Stimulation (flashing green)
Stimulation (flashing green)
LED indicators
The LED's indicate various conditions and activities:
All LED's are OFF =
battery is dead or removed, or device is OFF (switch position 0)
Upper green LED is flashing = system is operating and is in Stimulation Mode
Center yellow LED is flashing = system is running power-up diagnostics,
or system is operating and is in Rest Mode
Lower red LED is flashing =
system has a malfunction, requiring you to contact the supplier
Each time the Alleva-Wave P2 is powered on; the center yellow LED will flash three times. This is normal and indicates the
device is performing a self-diagnostic.
Coils
The magnetic field pulses are transmitted into the tissue between or adjacent to the two coils. Having two coils provides flexibility
in shaping and directing the magnetic fields. Configuring the coils properly is described later in this manual. During stimulation you
may hear a high-frequency clicking sound coming from the coils. This is normal and is caused by the rapid change of the magnetic
field produced by the coils. Listening for the clicking sound is a good way to assure the system is working properly.
Environmental Conditions that Affect Use
The Alleva-Wave system can be used during most normal daily activities. However, it cannot be used or worn when bathing,
swimming or during any activity that would expose the device to a wet environment. Do not submerge the device in water or any
liquid.
Caution
Do not expose the Alleva-Wave P2 system to a wet environment, or submerge the P2 system under
water, or clean the device using water. Doing so may permanently damage its electronics.