Stimulation and rest periods

 

 

Periods of rest are just as important as periods of stimulation to allow tissues to recover. The Alleva-Wave gives you two options 
to manage stimulation and rest periods. One option is for the Alleva-Wave to manage them for you. By moving the cycle profile 
switch to position 1 or 2, the P2 runs on an automatic and continuous 41 minute cycle, which includes a 30 minute stimulation 
period (Stimulation Mode) followed by an 11 minute rest period (Rest Mode). The 41 minute cycle repeats automatically; you do 
not have to do anything. This option is useful if you want to wear the device for many hours at a time without managing your 
stimulation and rest times.

 

 

The other option is for you to manage the periods of stimulation and rest. By moving the cycle profile switch to position 3, the 
Alleva-Wave runs continuously in Stimulation Mode with no rest periods. This option can be used when you want to have a brief 
and continuous stimulation period for perhaps an hour or two.

 

 

Two levels of magnetic field intensity

 

 

The Alleva-Wave can generate either standard intensity magnetic fields for mild tissue stimulation, or higher intensity fields for 
vigorous tissue stimulation. You can move the cycle profile switch to position 1 to generate standard intensity fields, or you can 
select position 2 or 3 to generate higher intensity fields.

 

 

When first using the Alleva-Wave it is recommended that you set the cycle profile to 1 (standard intensity with rest periods) on the 
first day or two as you become familiar with the device. Thereafter you can select cycle profile 2 (higher intensity with rest periods) 
or 3 (higher intensity with no rest periods).

 

 

Higher intensity fields are recommended when the injured tissue is far from the coils. It is also recommended after you feel 
comfortable with low intensity fields and you feel that more stimulation intensity would be beneficial for your tissue.

 

 

Cycle profile switch

 

 

The Alleva-Wave P2 operates on a 41 minute cycle that repeats continuously and automatically. Using the cycle profile switch, 
you can select from among four predefined cycle profiles.

 

 

Switch

 

Magnetic

 

Mode for

 

Mode for

 

position

 

field intensity

 

first 30 minutes

 

next 11 minutes

 

0

 

OFF

 

OFF

 

OFF

 

1

 

Standard

 

Stimulation (flashing green)

 

Rest (flashing yellow)

 

2

 

Higher

 

Stimulation (flashing green)

 

Rest (flashing yellow)

 

3

 

Higher

 

Stimulation (flashing green)

 

Stimulation (flashing green)

 

 

LED indicators

 

 

The LED's indicate various conditions and activities:

 

 

All LED's are OFF =

 

battery is dead or removed, or device is OFF (switch position 0)

 

Upper green LED is flashing =                  system is operating and is in Stimulation Mode

 

 

Center yellow LED is flashing =                system is running power-up diagnostics,

 

or system is operating and is in Rest Mode

 

 

Lower red LED is flashing =

 

system has a malfunction, requiring you to contact the supplier

 

 

Each time the Alleva-Wave P2 is powered on; the center yellow LED will flash three times. This is normal and indicates the 
device is performing a self-diagnostic.

 

 

Coils

 

 

The magnetic field pulses are transmitted into the tissue between or adjacent to the two coils. Having two coils provides flexibility 
in shaping and directing the magnetic fields. Configuring the coils properly is described later in this manual. During stimulation you 
may hear a high-frequency clicking sound coming from the coils. This is normal and is caused by the rapid change of the magnetic 
field produced by the coils. Listening for the clicking sound is a good way to assure the system is working properly.

 

 

Environmental Conditions that Affect Use

 

 

The Alleva-Wave  system can be used during most normal daily activities. However, it cannot be used or worn when bathing, 
swimming or during any activity that would expose the device to a wet environment. Do not submerge the device in water or any 
liquid.

 

 

Caution

 

 

Do not expose the Alleva-Wave P2 system to a wet environment, or submerge the P2 system under 
water, or clean the device using water. Doing so may permanently damage its electronics.

 

 

www.allevawave.com

 

4 of 12

 

P2User Instruction Manual

 

Summary of Contents for Alleva-Wave P2

Page 1: ...nformation 2 Warnings and Precautions 2 Purpose of the Alleva Wave P2 System 2 Description of the Alleva Wave P2 System 3 Environmental Conditions that Affect Use 4 System Setup 5 Operating Instructio...

Page 2: ...s If you feel pain during the use of the device you should stop using it and consult your physician If you experience increased pain or discomfort you should consider using a lower power setting if av...

Page 3: ...Mode during operation 3 2 Center yellow LED Indicates the system is running power up diagnostics before operation or is in Rest Mode during operation 2 Lower red LED Indicates there is a problem with...

Page 4: ...ile switch The Alleva Wave P2 operates on a 41 minute cycle that repeats continuously and automatically Using the cycle profile switch you can select from among four predefined cycle profiles Switch M...

Page 5: ...Insert a fully charged battery into the pulse generator Confirm that its terminals are correctly aligned to the positive and negative markings on the pulse generator A rechargeable NiMH battery will...

Page 6: ...thick parts of the body when the injury is relatively close to the surface within 1 to 2 inches 25 to 50 mm of the skin In these cases the coils are situated on the same side of the injury area adjac...

Page 7: ...ght later decide to place them obliquely at an angle To the extent possible it is best to try several options to determine what seems to work best The preferred placement of the coils might ultimately...

Page 8: ...e In both cases your body experiences longer term health gains as it s subjected to a variety of different external stimulations Another reason to change coil locations is the effect of accommodation...

Page 9: ...n their flat geometry However you are encouraged to flex and conform the coils to your body contours for both comfort and greater stimulation effectiveness Neck Shoulder Elbow Lower back www allevawav...

Page 10: ...ir daily routine and their response to the stimulation You are encouraged to experiment with usage patterns and coil locations to find what works best for you Also keeping a daily written record of yo...

Page 11: ...arther from the coils when I use the device Are longer coil cables available Or can I add an extension cable to the coil cable We offer only one length of coil cable We do this because the cable lengt...

Page 12: ...d I expect a reduction in pain Each person s response to the PEMF stimulation is different and we cannot guarantee results However pain reduction often occurs within the first 3 to 4 days of use and p...

Reviews: