Appendix | Industrial Process Interface for thermoMETER TIM 8 (Optional)
thermoMETER TIM 8 / thermoIMAGER TIM 40
i
Each stackable industrial PIF must have its own RS485 address. The address must be set directly on
the board and in the software. For the PIF which is the furthest away the 120R switch (TERM.- Termina-
tion) has to be set.
The process interface has an integrated fail-safe mode. This allows to control conditions like interruption of
cables, shut-down of the software etc. and to give out these conditions as an alarm.
Controlled conditions on camera
and software
Standard Process interface
TM-PIF-TIM8
Industrial Process interface
TM-PIFCBx-TIM8
Interruption USB cable to camera
Interruption data cable camera -
PIF
Interruption power supply PIF
Shut-down of TIM Connect
software
Crash of TIM Connect software
-
Fail-Safe-Output
0 V at analog output (AO)
Open contact
(fail-safe-relay)/ green LED off