Page 3
scanCONTROL 3000/BL / 3002/BL
5.
Connections, LED Displays
1 2
1 Multifunction port (power supply, I/O)
2 Ethernet port
Multifunction Port
Designation
Sensor
connector pin
Cable color
PCR3000-x
Notes
+Ub
9
Red
+ 11 V - 30 V DC
(rated value 24 V); max. 5 W
GND
2
Blue
0 V
+Laser on/off
3
White
Available with SI option
-Laser on/off
1
Brown
RS422
12
Red-blue
RS422
input or output
/RS422
11
Gray-pink
GND RS422
10
Purple
Ground connection RS422
In1
4
Green
Switching input In1
In2
6
Yellow
Switching input In2
In3
8
Gray
Switching input In3
In4
5
Pink
Switching input In4
GND-In
7
Black
Ground connection In
Shield
Housing
Black
Not electrically connected to GND
2
3
4
5
6
7
8
9
1
10
11
12
GND: Galvanically isolated from In1, In2, In3, In4,
RS422, laser on/off
Laser on/off: Input galvanically isolated from GND,
In1, In2, In3, In4, RS422
In1, In2, In3, In4, RS422: Input galvanically isolated
from GND and laser on/off
12-pin screw connector, view on solder side (cable)
RS422, Synchronization
The RS422 connection (pins 11 and 12 of the multifunction port) can be used in either
of the following configurations:
- RS422 (half-duplex): Load user modes, sensor control and transmit measurement
results (Modbus RTU or ASCII format).
- Synchronization/triggering: Synchronization or triggering using switching signals