Page 40
Measure
optoCONTROL 2600
Binary format
Data conversion
Start
0
0
6 Bit
(D5...D0)
Stop Start
0
1
6 Bit
(D11...D6)
Stop Start
1
0
6 Bit
(My...D12)
Stop
Fig. 43 Transmission format of a data word (example)
L-Byte
0
0
D5
D4
D3
D2
D1
D0
M-Byte 0
1
D11
D10
D9
D8
D7
D6
H-Byte 1
0
My
Mx
D15
D14
D13
D12
Fig. 44 Reception
My
Mx
Seg. 1
0
0
Seg. 2
0
1
Seg. 3
1
0
Seg. 4
1
1
Fig. 45 Measurement allocation for
Multi-segment
operating mode
D15
D14
D13
D12
D11
D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
Fig. 46 Result of conversion (digital value DW)
The following formula is used for the conversion of the digital values (DV) to the measurement values (MV in mm):
MV (mm) DV * 40.824 / 65519 - 0.4204872
The settings
Offset
and
Gain
for the display or the analog output have no effect on the digital value.
With errors a digital value of
≥
65520 is output.
Error messages during measurement output:
65521 No edge
65522 At the beginning of the picture
65523 At the end of the picture
65524 Dark - bright edge
65525 Bright - dark edge
65526 Min. number of edges
65527 Max. number of edges
65528 Invalid measuring program
65529 Segment 1st edge > 2nd edge
65530 Segment number of edges < last edge
65531 Invalid working distance
65533 Laser off
65534 Invalid float
65535 DMA setup error