Page 50
Operation
induSENSOR MSC 7210
Example B
Measuring range: 25 mm,
Start position X1 ... end position X2,
Output signal: 2 ... 10 VDC (4 ... 20 mA)
Set the frequency with J4, see Chap.
5.2.1
.
Set the phase with J2, see Chap.
5.2.1
.
Set the supply voltage for the sensor with J1, see
Chap.
5.2.2
.
S1 and S2 of J3 are „ON“, see
Fig. 12
Adjust the output to 2 VDC respectively 4 mA with
R1, see
Fig. 11
.
Put S1 and S2 of J3 „OFF“.
Move plunger to start position X1.
Adjust output to 2 VDC respectively 4 mA with R2.
Move plunger to end position X2.
Adjust gain with R3 until the output reaches
10 VDC respectively 20 mA.
10 V
2 V
0 mm
25 mm
X1
X2
Measuring range
i
The plunger respectively the sensor are fixed
to the target.