Operation
induSENSOR MSC 7210
5.2.3
Zero Point
Connect the sensor to the controller prior to calibration or measurement.
The electrical zero point can be shifted about ±70 % of the measuring range. Examples are shown for zero
point adjustment, see Chap.
5.3
.
5.2.4
Signal Gain
The signal gain can be changed about -20 % up to +270 % of the measuring range. Examples are shown for
gain adjustment, see Chap.
5.3
.
If the plunger is moved into the sensor, this results in an increase of the output signal, if the plunger is pulled
out further, this results in a decrease of the output signal.
Move the plunger by a defined range, in the ideal case by the complete linear measuring range and
adjust with R3 trim-pot the desired output voltage/current.
It makes no difference whether the plunger is pushed in or pulled out, except that this leads to a increase
respectively decrease of the output signal.