Page 49
Operation
induSENSOR MSC 7210
5.3
Examples
Example A
Measuring range: 25 mm,
Used measuring range: 25 mm,
Output signal: 2 ... 10 VDC (4 ... 20 mA)
The plunger is situated outside the range of the sen-
sor.
Set the frequency with J4, see Chap.
5.2.1
.
Set the phase with J2, see Chap.
5.2.1
.
Set the supply voltage for the sensor with J1, see
Chap.
5.2.2
.
Put S1 of J3 „OFF“ and S2 of J3 „ON“.
Adjust the output to 6.00 VDC respectively
12.00 mA (current output) with
R1, see
Fig. 11
.
Middleposition of the sensor.
Move the plunger into the centre of the sensor
until you get 6.00 VDC respectively 12.00 mA on
the output (electrical and mechanical output are
adjusted).
Core is in center position.
Move plunger about +12.5 mm.
Adjust output to 10.00 VDC respectively 20.00 mA
with R3.
10 V
2 V
0 mm
25 mm
Measuring range
i
The plunger respectively the sensor can be
adjusted mechanical during the settings.