FSC1/7 and FSC1000 Operation
FSC1/7, FSC1000
Depending on the setup configuration, press the button on the sen-
sor or keys 1 to 3 on the controller. The measurement will start
appr. 300 ms after you release the button.
Then enter the thickness of the calibration foil in the
Thickness
field in micrometers as an integer.
Fig. 16 Indicate the thickness of calibration foil 1
i
Select the distance between the support points in a range of max. 100 to 150 µm over the entire
measuring range. Make sure that the measuring area of the sensor, the calibration foil and the measur-
ing point are clean.
Place the second calibration foil on the substrate, place the
sensor on top, avoid vibrations/tilting and trigger the mea-
surement.
Calibration foil 2
Substrate
Depending on the setup configuration, press the button on the sen-
sor or keys 1 to 3 on the controller. The measurement will start 300
ms after you release the button.
Then enter the thickness of the calibration foil in the
Thickness
field in micrometers as an integer.
Fig. 17 Indicate the thickness of calibration foil