Changing the Lamp
Let the light source cool off for a few minutes before changing the lamp.
> Risk of burns by the still hot lamp
Do not touch the inside of the lamp (mirror, lamp bulb).
> Reducing the intensity and lifetime of the lamp by finger prints and dust
Switch off the
ON/OFF
switch, see
and disconnect the power supply cable from the wall socket.
Remove the screw, see
, on the front side of the light source (
1
), whereupon the right front door
with the lamp unit can be folded down (
2
).
Loosen the lamp out of the holder carefully (
3
), lift the spring clip slightly (
4
) and pull the lamp from the
lamp socket (
5
).
1
Remove the screw
2
Fold down
lamp unit
Holder
3
Remove
holder
CAUTION