Seite 27
Appendix| Change from SGS4311 to SGS4701
eddyNCDT SGS 4701
SGS4311
SGS4701
The controller is mounted close to the spin-
dle. Consider some space for it.
Components of the SGS4311 are not com-
patible with the SGS4701 series.
Voltage output 1.0 ... 9.0 V
for 150 ... 550 µm
It is necessary to adapt the output scaling
in the software/customer control system.
Voltage output 0.5 ... 9.5 V
for 100 ... 600 µm
The sensor cable length is permanently
defined for each system and must not be
changed.
Bandwidth 100 Hz
It is necessary to adapt the averaging
process in the software/customer control
system.
Bandwidth 2000 Hz
Cable length of max. 12 m at voltage
output
Optional accessories:
- Supply and output cable
PC4701-10 or
PC4701-10/ for use with drag cable
You can assemble a power and output
cable yourself using a conventional M12
connector.
Cable length > 12 m
Micro-Epsilon recommends to convert the
output signal from voltage to current.