INSTALLATION
4
INSTALLATION
Before installing the analyzer, read through this manual carefully and take note of all
warnings.
The XTC601-EX versions must only be installed by
suitably qualified personnel and in accordance with the
instructions provided and the terms of the applicable
product certificates.
Maintenance and servicing of the product must only be
carried out by suitably trained personnel or returned to an
approved Michell Instruments’ Service Center.
Before the cover is refitted, the flamepath/threaded
joint between the cover and body must be thoroughly
wiped clean of dirt, grit or other foreign substances and
then a thin coating of an approved form of non-setting
grease applied to joint/threads. Ensure O-ring is free from
damage.
Threaded covers must be screwed down until fully tight
(at least 7 full turns) and then locked in position with the
hex key provided.
The terms of the Hazardous Area Certification do not
permit oxygen measurement in ambient areas with
enriched oxygen
4.1
Unpacking
If sold separately (not part of a sampling system), the XTC601 will be supplied in a
custom box which should be retained for future use (such as service return). The box
contains a small carton containing 2 lid keys and 1 hex key (for the grub screw). Any
cable glands supplied will also be in the smaller carton.
Contents
:
•
XTC601 Binary Gas Analyzer
•
Test result sheet
•
Carton (containing: 2 x lid keys and 1 x hex key)
•
Cable entry options (if ordered)